Electrical connection
sensor connecting cable
Prosonic Flow 91
Connecting power sensor cables in the connection compartment
A0005843
Connecting the measuring system
a, b
Sensor connecting cables
c
Cover of cable gland
d
Rubber seal
e
Cable retaining sleeves
f
Ground disk
g
Cable gland holder
h
Seal
i
Cable holder
Supply voltage (power supply)
Transmitter:
• 85 to 260 V AC, 45 to 65 Hz
• 20 to 55 V AC, 45 to 65 Hz
• 16 to 62 V DC
Measuring sensors:
• Powered by the transmitter
Cable entry
Power supply and signal cables (inputs/outputs):
• Cable entry M20 x 1.5 (8 to 12 mm)
or
• Cable gland for cables with
∅
6 to 12 mm
• Threaded adapter ½" NPT, G ½"
Sensor connecting cable:
A special cable gland allows you to insert both sensor cables (per channel) into the connection compartment
simultaneously.
• Cable gland M20 x 1.5 for 2 x
∅
4 mm
or
• Threaded adapter ½" NPT, G ½"
a
b
a
b
h
a
b
A
c
d
e
f
g
i