Proline Promass P 300
50
Hauser
Shock resistance
Shock due to rough handling following IEC 60068-2-31
Interior cleaning
• Cleaning in place (CIP)
• Sterilization in place (SIP)
• Cleaning with pigs
Options
Oil- and grease-free version for wetted parts, without declaration
Order code for "Service", option HA
Electromagnetic
compatibility (EMC)
• As per IEC/EN 61326 and NAMUR Recommendation 21 (NE 21)
• Device version with PROFIBUS DP: Complies with emission limits for industry as per EN 50170
Volume 2, IEC 61784
The following applies for PROFIBUS DP: If baud rates > 1.5 MBaud, an EMC cable entry must be
used and the cable shield must continue as far as the terminal wherever possible.
Details are provided in the Declaration of Conformity.
Process
Medium temperature range
Standard version
–50 to +150 °C (–58 to +302 °F)
Order code for "Measuring tube
mat., wetted surface", option BB,
BC, BD
Extended temperature version
–50 to +205 °C (–58 to +401 °F)
Order code for "Measuring tube
mat., wetted surface", option TD, TG
Dependency of ambient temperature on medium temperature
T
a
T
m
B
A
A0031121
22
Exemplary representation, values in the table below.
T
a
Ambient temperature
T
m
Medium temperature
A
Maximum permitted medium temperature T
m
at T
a max
= 60 °C (140 °F); higher medium temperatures T
m
require a reduced ambient temperature T
a
B
Maximum permitted ambient temperature T
a
for the maximum specified medium temperature T
m
of the
sensor
Values for devices used in the hazardous area: