background image

The robust high

pressure flowmeter with a compact, easily accessible

transmitter

Application

• Measuring principle operates independently of physical fluid

properties such as viscosity or density

• For premium accuracy at highest process pressures, fully

suitable for offshore conditions

Device properties

• Measuring tube in 25Cr Duplex, 1.4410 (UNS S32750)

• Process pressure up to PN 250 (Class 1500)

• Nominal diameter: DN 80 to 250 (3 to 10")

• Compact dual-compartment housing with up to 3 I/Os

• Backlit display with touch control and WLAN access

• Remote display available

Your benefits

• Maximum safety – highest resistance to

stress corrosion cracking

• Fewer process measuring points – multivariable

measurement (flow, density, temperature)

• Space

saving installation – no inlet/outlet run needs

• Full access to process and diagnostic information –

numerous, freely combinable I/Os and fieldbuses

• Reduced complexity and variety – freely configurable I/O

functionality

• Integrated verification – Heartbeat Technology

Products

Solutions

Services

Technical Information

Proline Promass O 300

Coriolis flowmeter

TI01275D/06/EN/04.18

71415041

2018-11-01

Содержание Proline Promass O 300

Страница 1: ...using with up to 3 I Os Backlit display with touch control and WLAN access Remote display available Your benefits Maximum safety highest resistance to stress corrosion cracking Fewer process measuring...

Страница 2: ...s 48 Special mounting instructions 48 Environment 50 Ambient temperature range 50 Storage temperature 50 Climate class 50 Degree of protection 50 Vibration resistance 50 Shock resistance 50 Shock resi...

Страница 3: ...88 Concentration 89 Special density 89 Petroleum 89 OPC UA server 89 Accessories 89 Device specific accessories 89 Communication specific accessories 90 Service specific accessories 91 System componen...

Страница 4: ...nects the protectiv earth to the mains supply Outer ground terminal Connects the device to the plant grounding system Communication symbols Symbol Meaning Wireless Local Area Network WLAN Communicatio...

Страница 5: ...n the mass flow Instead of a constant rotational velocity the sensor uses oscillation In the sensor two parallel measuring tubes containing flowing fluid oscillate in antiphase acting like a tuning fo...

Страница 6: ...act version The transmitter and sensor form a mechanical unit Transmitter Promass 300 A0026708 Device versions and materials Transmitter housing Aluminum coated aluminum AlSi10Mg coated Cast stainless...

Страница 7: ...of specific functions to support protective measures on the operator s side These functions can be configured by the user and guarantee greater in operation safety if used correctly An overview of the...

Страница 8: ...extra is protected by the network key The WLAN authentication of the network key complies with the IEEE 802 11 standard When the device is delivered the network key is pre defined depending on the de...

Страница 9: ...and encrypted Access via service interface CDI RJ45 The device can be connected to a network via the service interface CDI RJ45 Device specific functions guarantee the secure operation of the device...

Страница 10: ...alculated with the formula below max G minimum max F G x G cG 2 di 2 3600 max G Maximum full scale value for gas kg h max F Maximum full scale value for liquid kg h max G max F max G can never be grea...

Страница 11: ...ers and temperature measuring devices can be ordered from Endress Hauser see Accessories section 92 It is recommended to read in external measured values to calculate the corrected volume flow HART pr...

Страница 12: ...t values DC 3 to 30 V If status input is active ON Ri 3 k Response time Adjustable 5 to 200 ms Input signal level Low signal DC 3 to 5 V High signal DC 12 to 30 V Assignable functions Off Reset the in...

Страница 13: ...rder code for Output input 2 021 Not assigned A A A A A A A A A A Current output 0 4 to 20 mA B B B B B B B Current output 0 4 to 20 mA Ex i C C C User configurable input output 1 D D D D D D D Pulse...

Страница 14: ...Oscillation damping 0 Signal asymmetry Exciter current 0 The range of options increases if the measuring device has one or more application packages PROFIBUS PA PROFIBUS PA In accordance with EN 5017...

Страница 15: ...ssive Load 0 to 700 Resolution 0 38 A Damping Adjustable 0 07 to 999 s Assignable measured variables Mass flow Volume flow Corrected volume flow Density Reference density Temperature Electronic temper...

Страница 16: ...low Density Reference density Temperature Electronic temperature Oscillation frequency 0 Oscillation damping 0 Signal asymmetry Exciter current 0 The range of options increases if the measuring device...

Страница 17: ...ior Can be set to NO normally open factory setting NC normally closed Maximum switching capacity passive DC 30 V 0 1 A AC 30 V 0 5 A Assignable functions Off On Diagnostic behavior Limit value Mass fl...

Страница 18: ...ance with PROFIBUS PA Profile 3 02 EtherNet IP Device diagnostics Device condition can be read out in Input Assembly PROFINET Device diagnostics According to Application Layer protocol for decentraliz...

Страница 19: ...us Open Closed Relay output Failure mode Choose from Current status Open Closed Local display Plain text display With information on cause and remedial measures Backlight Red backlighting indicates a...

Страница 20: ...7 Option BA Current output 4 to 20 mA HART UN 30 VDC UM 250 VAC Option GA PROFIBUS PA UN 30 VDC UM 250 VAC Option LA PROFIBUS DP UN 30 VDC UM 250 VAC Option MA Modbus RS485 UN 30 VDC UM 250 VAC Option...

Страница 21: ...ersion transmitter Order code for Output input 2 Output input 3 Output type Intrinsically safe values or NIFW values Output input 2 Output input 3 24 25 22 23 Option C Current output 4 to 20 mA Ex i U...

Страница 22: ...D file Earlier models Promass 80 PROFIBUS PA ID No 1528 hex Extended GSD file EH3x1528 gsd Standard GSD file EH3_1528 gsd Promass 83 PROFIBUS PA ID No 152A hex Extended GSD file EH3x152A gsd Standard...

Страница 23: ...ity Auto polarity for automatic correction of crossed TxD and RxD pairs Supported CIP connections Max 3 connections Explicit connections Max 6 connections I O connections Max 6 connections scanner Con...

Страница 24: ...ufacturer specific software FieldCare DeviceCare Web browser Device master file GSD can be read out via the integrated Web server of the measuring device Configuration of the device name DIP switches...

Страница 25: ...vice Link Capabilities Slot time 4 Min delay between PDU 8 Max response delay 16 System integration Information regarding system integration Operating Instructions 92 Cyclic data transmission Descript...

Страница 26: ...on system integration Operating Instructions 92 Modbus RS485 information Function codes Register information Response time Modbus data map Power supply Terminal assignment Transmitter supply voltage i...

Страница 27: ...device version ordered 13 Terminal assignment of the remote display and operating module 32 Device plugs available Device plugs may not be used in hazardous areas Device plugs for fieldbus systems Or...

Страница 28: ...ctrical connection 2 3 L N P U Connector M12 1 R 1 2 S 1 2 T 1 2 V 1 2 Connector M12 1 Connector M12 1 1 Cannot be combined with an external WLAN antenna order code for Enclosed accessories option P8...

Страница 29: ...4 Phoenix part no 1543223 SACC M12MSD 4Q When using the device in a hazardous location use a suitably certified plug Service interface Order code for Accessories mounted option NB Adapter RJ45 M12 ser...

Страница 30: ...tion NE21 Current consumption Transmitter Max 400 mA 24 V Max 200 mA 110 V 50 60 Hz 230 V 50 60 Hz Power supply failure Totalizers stop at the last value measured Depending on the device version the c...

Страница 31: ...to an M12 connector mounted in the cable entry Therefore the connection to the service interface can be established via an M12 connector without opening the device Network connection via service inter...

Страница 32: ...e Only one display or operation unit may be connected to the transmitter at any one time 1 3 81 Vcc 82 Gnd 83 A 84 B 81 Vcc 82 Gnd 83 A 84 B 2 4 5 A0027518 1 Remote display and operating module DKX001...

Страница 33: ...ons 40 4 Analog display unit observe maximum load 14 5 Transmitter HART input 2 4 20 mA 4 1 2 3 3 6 5 A0028763 4 Connection example for HART input with a common negative passive 1 Automation system wi...

Страница 34: ...x 5 Measuring device 6 Local grounding 7 Bus terminator 8 Potential matching line PROFIBUS DP 2 1 A B 3 4 4 A B A B A0028765 6 Connection example for PROFIBUS DP non hazardous area and Zone 2 Div 2 1...

Страница 35: ...Device plug 5 Transmitter EtherNet IP DLR Device Level Ring 1 2 4 5 5 3 A0027544 1 Control system e g PLC 2 Ethernet switch 3 Observe cable specifications 40 4 Connecting cable between the two transmi...

Страница 36: ...two transmitters 5 Transmitter FOUNDATION Fieldbus 2 1 3 4 7 8 6 6 6 6 5 6 6 5 A0028768 9 Connection example for FOUNDATION Fieldbus 1 Control system e g PLC 2 Power Conditioner FOUNDATION Fieldbus 3...

Страница 37: ...rve cable specifications 3 Distribution box 4 Transmitter Current output 4 20 mA 4 20 mA 2 1 3 A0028758 11 Connection example for 4 20 mA current output active 1 Automation system with current input e...

Страница 38: ...nput values 15 Switch output 1 2 3 A0028760 14 Connection example for switch output passive 1 Automation system with switch input e g PLC 2 Power supply 3 Transmitter Observe input values 15 Double pu...

Страница 39: ...se output 5 Double pulse output slave phase shifted Relay output 1 2 3 A0028760 17 Connection example for relay output passive 1 Automation system with relay input e g PLC 2 Power supply 3 Transmitter...

Страница 40: ...vailable for certain device versions 27 Cable specification Permitted temperature range The installation guidelines that apply in the country of installation must be observed The cables must be suitab...

Страница 41: ...of ODVA Organization PROFINET Standard IEC 61156 6 specifies CAT 5 as the minimum category for a cable used for PROFINET CAT 5e and CAT 6 are recommended For more information on planning and installi...

Страница 42: ...n 1 Cable length Maximum 300 m 1000 ft see the following table Cross section Cable length for use in Non hazardous area Hazardous area Zone 2 Class I Division 2 Hazardous area Zone 1 Class I Division...

Страница 43: ...ion flow option D for mass flow 0 10 o r Mass flow gases 0 35 o r Density liquids Under reference operating conditions Standard density calibration 1 Wide range Density specification 2 3 g cm g cm g c...

Страница 44: ...9400 2940 1470 588 294 58 80 10 80850 8085 4043 1617 808 5 161 7 Accuracy of outputs The outputs have the following base accuracy specifications Current output Accuracy 5 A Pulse frequency output o r...

Страница 45: ...or is 0 00005 g cm3 C 0 000025 g cm3 F Field density calibration is possible Wide range density specification special density calibration If the process temperature is outside the valid range 43 the m...

Страница 46: ...21332 BaseAccu A0021339 ZeroPoint BaseAccu 100 A0021333 ZeroPoint MeasValue 100 A0021334 Calculation of the maximum repeatability as a function of the flow rate Flow rate Maximum repeatability in o r...

Страница 47: ...pipeline Pipe restrictions or the use of an orifice with a smaller cross section than the nominal diameter prevent the sensor running empty while measurement is in progress 1 2 3 4 5 A0028773 20 Inst...

Страница 48: ...To maintain the maximum ambient temperature for the transmitter this orientation is recommended If a sensor is installed horizontally with a curved measuring tube match the position of the sensor to t...

Страница 49: ...ocation for the rupture disk In the event of a failure of the rupture disk a discharge device can be screwed onto the internal thread of the rupture disk in order to drain off any escaping medium RUPT...

Страница 50: ...warm climatic regions You can order a weather protection cover from Endress Hauser 89 Storage temperature 50 to 80 C 58 to 176 F Climate class DIN EN 60068 2 38 test Z AD Degree of protection Measuri...

Страница 51: ...ld must continue as far as the terminal wherever possible Details are provided in the Declaration of Conformity Process Medium temperature range 40 to 205 C 40 to 401 F Dependency of ambient temperatu...

Страница 52: ...imum permissible medium pressure depending on the specific medium temperature Flange connection according to EN 1092 1 DIN 2501 300 F 350 250 200 150 100 50 0 50 C 200 150 100 50 0 50 400 120 140 160...

Страница 53: ...or housing is filled with dry nitrogen gas and protects the electronics and mechanics inside If a measuring tube fails e g due to process characteristics like corrosive or abrasive fluids the fluid wi...

Страница 54: ...corresponding type test declaration can be ordered with the device order code for Additional approval option LN Sensor housing burst pressure type test Pressure rating according to ASME BPVC DN Senso...

Страница 55: ...transmitter housing 80 C 176 F Thermal insulation with extended neck free We recommend that you do not insulate the extended neck in order to ensure optimum dissipation of heat A0034391 26 Thermal in...

Страница 56: ...ted fluid temperature in these applications is 25 to 55 C 13 to 131 F The device is used with a legally controlled totalizer on the local display and optionally with legally controlled outputs Measuri...

Страница 57: ...y values 30 mm 3 Dependent on respective process connection Order code for Housing option A Aluminum coated Ex d DN A 1 B C D E F G 2 H I 2 K L M mm mm mm mm mm mm mm mm mm mm mm mm mm 80 188 85 103 2...

Страница 58: ...E F G H I K L M mm mm mm mm mm mm mm mm mm mm mm mm mm 150 186 85 101 378 433 811 221 63 158 68 9 2 218 250 186 85 101 548 461 1009 221 63 158 99 1 2 305 1 Depending on the cable gland used values up...

Страница 59: ...1092 1 Form B1 DIN 2501 PN40 Stainless steel 1 4410 F53 25Cr Duplex Super Duplex Order code for Process connection option DFD DN mm A mm B mm C mm D mm E mm L mm 250 450 385 12 33 38 258 8 1844 Flange...

Страница 60: ...5 7 1476 Flange according to EN 1092 1 Form B2 DIN 2501 PN250 Stainless steel 1 4410 F53 25Cr Duplex Super Duplex Order code for Process connection option DBD DN mm A mm B mm C mm D mm E mm L mm 80 25...

Страница 61: ...A mm B mm C mm D mm E mm L mm 80 240 190 5 8 25 4 45 1 78 0 962 100 290 235 8 31 8 51 4 102 4 1251 150 380 317 5 12 31 8 62 6 154 1 1513 250 545 469 9 12 38 1 76 9 254 5 2016 Flange according to ASME...

Страница 62: ...Super Duplex Order code for Process connection option NDD DN mm A mm B mm C mm D mm E mm L mm 250 400 355 12 25 24 250 1774 Flange JIS B2220 20K Stainless steel 1 4410 F53 25Cr Duplex Super Duplex Ord...

Страница 63: ...ve cover 146 5 75 48 1 9 12 0 47 280 11 0 255 10 0 134 5 3 30 1 18 A0029553 Remote display and operating module DKX001 136 5 35 114 4 49 86 3 39 78 3 07 103 4 06 12 0 47 7 0 0 27 A0028921 28 Engineeri...

Страница 64: ...in External WLAN antenna mounted with cable The external WLAN antenna can be mounted separately from the transmitter if the transmission reception conditions at the transmitter mounting location are p...

Страница 65: ...n in in in in 3 7 40 3 35 4 06 7 87 15 63 23 5 8 54 2 28 5 83 1 59 3 5 47 4 7 40 3 35 4 06 10 16 26 26 26 8 54 2 28 5 83 2 02 3 6 93 6 7 40 3 35 4 06 14 88 17 05 31 93 8 54 2 28 5 83 2 71 3 8 58 10 7...

Страница 66: ...uper Duplex Order code for Process connection option ABD DN in A in B in C in D in E in L in 10 17 52 15 25 16 1 13 1 89 10 02 73 31 Flange according to ASME B16 5 Class 600 Schedule 40 Stainless stee...

Страница 67: ...dule 40 Stainless steel 1 4410 F53 25Cr Duplex Super Duplex Order code for Process connection option AED DN in A in B in C in D in E in L in 3 9 45 7 5 8 1 0 1 81 3 07 37 91 4 11 42 9 25 8 1 25 2 06 4...

Страница 68: ...1 2 1 1 2 A0028914 1 Connection nipple for purge connections order code for Sensor options option CH Purge connection 2 Connection nipple with rupture disk order code for Sensor option option CA Ruptu...

Страница 69: ...WLAN antenna The external WLAN antenna is not suitable for use in hygienic applications External WLAN antenna mounted on device 68 2 7 105 4 1 173 6 8 A0028923 32 Engineering unit mm in External WLAN...

Страница 70: ...r the hazardous area Order code for Housing option A Aluminum coated Ex d 2 kg 4 4 lbs Cast transmitter version stainless Order code for Housing option L Cast stainless 6 kg 13 lbs Weight in SI units...

Страница 71: ...ter for cable entry with internal thread NPT Order code for Housing option L Cast stainless The various cable entries are suitable for hazardous and non hazardous areas Cable entry cable gland Materia...

Страница 72: ...cess via Web server or SmartBlue app 91 WLAN access to the device via mobile handheld terminal tablet or smart phone Reliable operation Operation in local language 72 Uniform operating philosophy appl...

Страница 73: ...ting elements External operation via touch control 3 optical keys without opening the housing Operating elements also accessible in the various zones of the hazardous area Via remote display and opera...

Страница 74: ...code for Electrical connection Connecting cable 42 Dimensions 63 Remote operation Via HART protocol This communication interface is available in device versions with a HART output 1 2 3 5 8 4 7 6 A00...

Страница 75: ...S Device Manager SIMATIC PDM with COM DTM CDI Communication TCP IP 6 Commubox FXA195 USB 7 Field Xpert SFX350 or SFX370 8 Field Xpert SMT70 9 VIATOR Bluetooth modem with connecting cable 10 Transmitte...

Страница 76: ...DP network 4 Measuring device Via PROFIBUS PA network This communication interface is available in device versions with PROFIBUS PA 2 3 1 7 6 5 4 A0028838 41 Options for remote operation via PROFIBUS...

Страница 77: ...topology 4 2 1 3 5 5 5 A0032078 43 Options for remote operation via EtherNet IP network star topology 1 Automation system e g RSLogix Rockwell Automation 2 Workstation for measuring device operation...

Страница 78: ...eviceCare with COM DTM CDI Communication TCP IP 4 Ethernet switch 5 Measuring device Via PROFINET network This communication interface is available in device versions with PROFINET Star topology 3 1 2...

Страница 79: ...lly available Order code for Accessories option NB Adapter RJ45 M12 service interface The adapter connects the service interface CDI RJ45 to an M12 connector mounted in the cable entry Therefore the c...

Страница 80: ...to 11 Degree of protection IP67 Available antennas Internal antenna External antenna optional In the event of poor transmission reception conditions at the place of installation Available as an acces...

Страница 81: ...he OPC UA Server application package 93 Supported operating tools Different operating tools can be used for local or remote access to the measuring device Depending on the operating tool used access i...

Страница 82: ...device is also displayed and allows the user to monitor the status of the device Furthermore the device data can be managed and the network parameters can be configured A device that has a WLAN inter...

Страница 83: ...easuring device is ready for operation again immediately without any errors If exchanging the electronics module e g I O electronics module Once the electronics module has been replaced the software o...

Страница 84: ...the EMC requirements of the Australian Communications and Media Authority ACMA Ex approval The measuring device is certified for use in hazardous areas and the relevant safety instructions are provid...

Страница 85: ...ure order code for Additional approval option LA and SIL 3 multichannel architecture with homogeneous redundancy and is independently evaluated and certified by the T V in accordance with IEC 61508 Th...

Страница 86: ...on the sensor nameplate Endress Hauser confirms conformity with the Essential Safety Requirements specified in Appendix I of the Pressure Equipment Directive 2014 68 EU Devices bearing this marking P...

Страница 87: ...bility EMC requirements NAMUR NE 21 Electromagnetic compatibility EMC of industrial process and laboratory control equipment NAMUR NE 32 Data retention in the event of a power failure in field and con...

Страница 88: ...escription Extended HistoROM Comprises extended functions concerning the event log and the activation of the measured value memory Event log Memory volume is extended from 20 message entries standard...

Страница 89: ...important parameters for the Oil Gas Industry can be calculated and displayed with this application package Corrected volume flow and calculated reference density in accordance with the API Manual of...

Страница 90: ...N antenna is not suitable for use in hygienic applications Further information on the WLAN interface 79 Order number 71351317 Installation Instructions EA01238D Protective cover Is used to protect the...

Страница 91: ...pplicator Software for selecting and sizing Endress Hauser measuring devices Choice of measuring devices for industrial requirements Calculation of all the necessary data for identifying the optimum f...

Страница 92: ...ture transmitters can be used in all applications and are suitable for the measurement of gases steam and liquids They can be used to read in the medium temperature Fields of Activity document FA00006...

Страница 93: ...nA XA01507D INMETRO Ex d Ex de XA01468D INMETRO Ex ec XA01470D NEPSI Ex d Ex de XA01469D NEPSI Ex nA XA01471D Remote display and operating module DKX001 Contents Documentation code ATEX IECEx Ex i XA0...

Страница 94: ...r SD01688D SD01689D Installation Instructions Contents Comment Installation instructions for spare part sets and accessories Documentation code specified for each individual accessory Registered trade...

Страница 95: ......

Страница 96: ...www addresses endress com...

Отзывы: