16.6 Performance characteristics
Reference operating
conditions
• Error limits based on ISO 11631
• Water with +15 to +45 °C (+59 to +113 °F) at2 to 6 bar (29 to 87 psi)
• Specifications as per calibration protocol
• Accuracy based on accredited calibration rigs that are traced to ISO 17025.
To obtain measured errors, use the
Applicator
sizing tool (→ 111)
Maximum measured error
o.r. = of reading; 1 g/cm³ = 1 kg/l; T = medium temperature
Base accuracy
Mass flow and volume flow (liquids)
±0.05 % o.r. (PremiumCal, for mass flow)
±0.10 %
Mass flow (gases)
±0.35 % o.r.
Density (liquids)
• Reference conditions:±0.0005 g/cm³
• Standard density calibration:±0.01 g/cm³
(valid over the entire temperature range and density range )
• Wide-range density specification (order code for "Application package", option EF "Special
density and concentration ") : ±0.001 g/cm³ (valid range for special density calibration:
0 to 2 g/cm³, +5 to +80 °C (+41 to +176 °F))
Temperature
±0.5 °C ± 0.005 · T °C (±0.9 °F ± 0.003 · (T – 32) °F)
Zero point stability
DN
Zero point stability
[mm]
[in]
[kg/h]
[lb/min]
80
3
9.0
0.330
100
4
14.0
0.514
150
6
32.0
1.17
Flow values
Flow values as turndown parameter depending on nominal diameter.
SI units
DN
1:1
1:10
1:20
1:50
1:100
1:500
[mm]
[kg/h]
[kg/h]
[kg/h]
[kg/h]
[kg/h]
[kg/h]
80
180
000
18
000
9
000
3
600
1
800
360
100
350
000
35
000
17
500
7
000
3
500
700
150
800
000
80
000
40
000
16
000
8
000
1
600