Proline Promass 40E
8
Hauser
US units
Accuracy of outputs
o.r. = of reading; o.f.s. = of full scale value
The output accuracy must be factored into the measured error if analog outputs are used, but can be
ignored for fieldbus outputs (e.g. Modbus RS485, EtherNet/IP).
Current output
Accuracy: Max. ±0.05 % o.f.s. or ±5
μ
A
Pulse/frequency output
Accuracy: Max. ±50 % ppm o.r.
Repeatability
Design fundamentals
→
9
o.r. = of reading; 1 g/cm
3
= 1 kg/l; T = fluid temperature
Base repeatability
Mass flow and volume flow (liquids)
±0.25% o.r.
Mass flow (gases)
±0.50% o.r.
Temperature
±0.25 °C ± 0.0025 · T °C (±0.45 °F ± 0.0015 · (T–32) °F)
Response time
• The response time depends on the configuration (damping).
• Response time in the event of erratic changes in the measured variable (only mass flow): after
100 ms 95 % of the full scale value.
Influence of fluid
temperature
When there is a difference between the temperature for zero point adjustment and the process
temperature, the typical measured error of the Promass sensor is ±0.0003% of the full scale value / °C
(±0.0001% of the full scale value / °F).
DN
1:1
1:10
1:20
1:50
1:100
1:500
[in]
[lb/min]
[lb/min]
[lb/min]
[lb/min]
[lb/min]
[lb/min]
³
⁄
₈
73.50
7.350
3.675
1.470
0.735
0.147
½
238.9
23.89
11.95
4.778
2.389
0.478
1
661.5
66.15
33.08
13.23
6.615
1.323
1½
1654
165.4
82.70
33.08
16.54
3.308
2
2573
257.3
128.7
51.46
25.73
5.146
3
6615
661.5
330.8
132.3
66.15
13.23