10.5.4 Performing electrode cleaning
The
Electrode cleaning circuit
wizard guides the user systematically through all the
parameters that have to be set for configuring electrode cleaning.
The wizard only appears if the device was ordered with an electrode cleaning circuit.
Navigation
"Setup" menu → Advanced setup → Electrode cleaning circuit
‣
Electrode cleaning circuit
Electrode cleaning circuit
ECC duration
ECC recovery time
ECC cleaning cycle
ECC Polarity
Parameter overview with brief description
Parameter
Prerequisite
Description
Selection / User
entry / User
interface
Factory setting
Electrode cleaning circuit
For the following order code:
"Application package", option
EC
"ECC electrode cleaning"
Enable the cyclic electrode
cleaning circuit.
• Off
• On
Off
ECC duration
For the following order code:
"Application package", option
EC
"ECC electrode cleaning"
Enter the duration of electrode
cleaning in seconds.
0.01 to 30 s
2 s
ECC recovery time
For the following order code:
"Application package", option
EC
"ECC electrode cleaning"
Define recovery time after
electrode cleaning. During this
time the current output values
will be held at last valid value.
1 to 600 s
5 s
ECC cleaning cycle
For the following order code:
"Application package", option
EC
"ECC electrode cleaning"
Enter the pause duration
between electrode cleaning
cycles.
0.5 to 168 h
0.66 h
ECC Polarity
For the following order code:
"Application package", option
EC
"ECC electrode cleaning"
Select the polarity of the
electrode cleaning circuit.
• Positive
• Negative
Depends on the
electrode material:
• Platinum:
Negative
option
• Tantalum, Alloy
C22, stainless
steel:
Positive
option
10.5.5 WLAN configuration
The
WLAN Settings
submenu guides the user systematically through all the parameters
that have to be set for the WLAN configuration.