12.7.4 Diagnostic of process
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
803 Current loop
1. Check wiring
2. Change I/O module
–
Measured variable status
Quality
Bad
Quality substatus
Process related
Coding (hex)
0x28 to 0x2B
Status signal
F
Diagnostic behavior
Alarm
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
832 Electronic temperature too high
Reduce ambient temperature
• Conductivity
• Corrected conductivity
• Measured values 1
• Measured values 2
• Measured values 3
• Density
• Electronic temperature
• Empty pipe detection
• Flow velocity
• Low flow cut off
• Mass flow
• Reference density
• Corrected volume flow
• Temperature
• Volume flow
Measured variable status [from the factory]
Quality
Bad
Quality substatus
Process related
Coding (hex)
0x28 to 0x2B
Status signal
S
Diagnostic behavior
Warning
1)
Diagnostic behavior can be changed. This causes the overall status of the measured variable to change.
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
833 Electronic temperature too low
Increase ambient temperature
• Conductivity
• Corrected conductivity
• Measured values 1
• Measured values 2
• Measured values 3
• Density
• Electronic temperature
• Empty pipe detection
• Flow velocity
• Low flow cut off
• Mass flow
• Reference density
• Corrected volume flow
• Temperature
• Volume flow
Measured variable status [from the factory]
Quality
Bad
Quality substatus
Process related
Coding (hex)
0x28 to 0x2B
Status signal
S
Diagnostic behavior
Warning
1)
Diagnostic behavior can be changed. This causes the overall status of the measured variable to change.