Proline Promag H 500 PROFIBUS DP
Commissioning
Hauser
109
Parameter
Prerequisite
Description
Selection
Factory setting
Conductivity unit
The
On
option is selected in
the
Conductivity
measurement
parameter
parameter.
Select conductivity unit.
Effect
The selected unit applies for:
Simulation process variable
Unit choose list
µS/cm
Temperature unit
–
Select temperature unit.
Result
The selected unit applies for:
•
Temperature
parameter
•
Maximum value
parameter
•
Minimum value
parameter
•
External temperature
parameter
•
Maximum value
parameter
•
Minimum value
parameter
Unit choose list
Country-specific:
• °C
• °F
Mass flow unit
–
Select mass flow unit.
Result
The selected unit applies for:
• Output
• Low flow cut off
• Simulation process variable
Unit choose list
Country-specific:
• kg/h
• lb/min
Mass unit
–
Select mass unit.
Unit choose list
Country-specific:
• kg
• lb
Density unit
–
Select density unit.
Result
The selected unit applies for:
• Output
• Simulation process variable
Unit choose list
Country-specific:
• kg/l
• lb/ft³
Corrected volume flow unit
–
Select corrected volume flow
unit.
Result
The selected unit applies for:
Corrected volume flow
parameter (→ 149)
Unit choose list
Country-specific:
• Nl/h
• Sft³/h
Corrected volume unit
–
Select corrected volume unit.
Unit choose list
Country-specific:
• Nm³
• Sft³
10.6.3 Configuring communication interface
The
Communication
submenu guides you systematically through all the parameters that
have to be configured for selecting and setting the communication interface.
Navigation
"Setup" menu → Communication
‣
Communication
Device address