Caution!
Do not return a measuring device if you are not absolutely certain that all traces of hazardous
substances have been removed, e.g. substances which have penetrated crevices or diffused through
plastic.
Costs incurred for waste disposal and injury (burns, etc.) due to inadequate cleaning will be charged
to the owner-operator.
The following steps must be taken before returning a flow measuring device to Hauser,
e.g. for repair or calibration:
• Always enclose a duly completed "Declaration of contamination" form. Only then can
Hauser transport, examine and repair a returned device.
• Enclose special handling instructions if necessary, for example a safety data sheet as per
EC REACH Regulation No. 1907/2006.
• Remove all residues. Pay special attention to the grooves for seals and crevices which could
contain residues. This is particularly important if the substance is hazardous to health, e.g.
flammable, toxic, caustic, carcinogenic, etc.
!
Note!
You will find a preprinted "Declaration of contamination" form at the back of these Operating
Instructions.
9.6
Disposal
Please observe the regulations applicable in your country or region.
9.7
Software history
Date
Software-Version
Changes to software
Documentation
10.2009
3.00.XX
Introduction of new FOUNDATION Fieldbus I/O
board
Reduced execution times:
Analog Input function blocks 1 to 5 (18 ms)
PID function block (25 ms)
Discrete Output function block (18 ms)
Integrator function block (18 ms)
Software adjustments:
- ITK version: 5.01
- CFF version: 1.8
71089883/07.09
01.2007
2.00.XX
New execution times:
• Analog Input function blocks 1 to 5 (20 ms)
• Discrete Output function block (20 ms)
• PID function block (50 ms)
New function blocks:
• Arithmetic function block (20 ms)
• Input Selector function block (20 ms)
• Signal Characterizer function block (20 ms)
• Integrator function block (25 ms)
Method:
• Communication
• Commissioning
ITK Version: 5.01
71031357/09.06
09.2006
1.01.02
Original software