4 Wiring
Proline Promag 51
50
Hauser
4.4
Degree of protection
The devices fulfill all the requirements for IP 67. Compliance with the following points is mandatory
following installation in the field or servicing, in order to ensure that IP 67 protection is maintained:
• The housing seals must be clean and undamaged when inserted into their grooves. The seals must
be dried, cleaned or replaced if necessary.
• All threaded fasteners and screw covers must be firmly tightened.
• The cables used for connection must be of the specified outside diameter (see Page 107).
• Firmly tighten the cable entries (Fig. 38).
• The cables must loop down before they enter the cable entries (“water trap”, Fig. 38). This
arrangement prevents moisture penetrating the entry. Always install the
measuring device in such a way that the cable entries do not point up.
• Remove all unused cable entries and insert plugs instead.
• Do not remove the grommet from the cable entry.
F06-xxxxxxxx-04-xx-xx-xx-005
Fig. 38:
Installation instructions, cable entries
"
Caution!
Do not loosen the threaded fasteners of the Promag sensor housing, as otherwise the degree of
protection guaranteed by Hauser no longer applies.
!
Note!
The Promag W and Promag P sensors can be supplied with IP 68 rating (permanent immersion in
water to a depth of 3 meters). In this case the transmitter must be installed remote from the sensor.
Содержание Proline Promag 51
Страница 4: ...QUICK SETUP commissioning Proline Promag 51 4 Endress Hauser...
Страница 10: ...1 Safety instructions Proline Promag 51 10 Endress Hauser...
Страница 52: ...4 Wiring Proline Promag 51 52 Endress Hauser...
Страница 80: ...7 Maintenance Proline Promag 51 80 Endress Hauser...
Страница 104: ...9 Trouble shooting Proline Promag 51 104 Endress Hauser...
Страница 122: ...11 Index Proline Promag 51 122 Endress Hauser...
Страница 124: ...www endress com worldwide BA080D 06 en 03 05 50101924 FM SGML6 0...