Proline Promag 300 Modbus RS485
Description of device parameters
Hauser
93
Additional information
Description
Corrected value = (factor × value) + offset
Temp. offset
Navigation
Expert → Sensor → Sensor adjustm. → Variable adjust → Temp. offset (1868)
Prerequisite
One of the following conditions is met:
• Order code for "Sensor option", option
CI
"Medium temperature measurement"
or
• The temperature is read into the flowmeter from an external device.
Description
Use this function to enter the zero point shift for the temperature trim. The temperature
unit on which the shift is based is 1 K.
User entry
Signed floating-point number
Factory setting
0 K
Additional information
Description
Corrected value = (factor × value) + offset
Temp. factor
Navigation
Expert → Sensor → Sensor adjustm. → Variable adjust → Temp. factor (1869)
Prerequisite
One of the following conditions is met:
• Order code for "Sensor option", option
CI
"Medium temperature measurement"
or
• The temperature is read into the flowmeter from an external device.
Description
Use this function to enter a quantity factor (without time) for the temperature. This
multiplication factor is applied over the temperature range.
User entry
Positive floating-point number
Factory setting
1
Additional information
Description
Corrected value = (factor × value) + offset