Description of device parameters
Proline Promag 300 Modbus RS485
90
Hauser
Mass flow factor (1846)
Conduct. offset (1848)
Conduct. factor (1849)
Corr. vol offset (1866)
Corr. vol factor (1867)
Temp. offset (1868)
Temp. factor (1869)
Corr.cond.offset (1870)
Corr.cond.factor (1871)
Flow vel. offset (1879)
Flow vel. factor (1880)
Vol. flow offset
Navigation
Expert → Sensor → Sensor adjustm. → Variable adjust → Vol. flow offset (1831)
Description
Use this function to enter the zero point shift for the volume flow trim. The volume flow
unit on which the shift is based is m³/s.
User entry
Signed floating-point number
Factory setting
0 m³/s
Additional information
Description
Corrected value = (factor × value) + offset
Vol. flow factor
Navigation
Expert → Sensor → Sensor adjustm. → Variable adjust → Vol. flow factor (1832)
User entry
Positive floating-point number
Factory setting
1