Flow vel. factor
Navigation
Expert → Sensor → Sensor adjustm. → Variable adjust → Flow vel. factor (1880)
Description
Use this function to enter a quantity factor (without time) for the flow velocity. This
multiplication factor is applied over the flow velocity range.
User entry
Positive floating-point number
Factory setting
1
Additional information
Description
Corrected value = (factor × value) + offset
3.2.6
"Calibration" submenu
Navigation
Expert → Sensor → Calibration
‣
Calibration
Nominal diameter (2807)
Cal. factor (6522)
Zero point (6546)
Cond. cal. fact. (6718)
Nominal diameter
Navigation
Expert → Sensor → Calibration → Nominal diameter (2807)
Description
Displays the nominal diameter of the sensor.
User interface
DNxx / x"
Factory setting
Depends on the size of the sensor
Additional information
Description
The value is also specified on the sensor nameplate.