Description of device parameters
Proline Promag 300 Modbus RS485
194
Hauser
Additional information
Description
This defines the interval between the individual data points in the data log, and thus the
maximum loggable process time T
log
:
• If 1 logging channel is used: T
log
= 1000 × t
log
• If 2 logging channels are used: T
log
= 500 × t
log
• If 3 logging channels are used: T
log
= 333 × t
log
• If 4 logging channels are used: T
log
= 250 × t
log
Once this time elapses, the oldest data points in the data log are cyclically overwritten such
that a time of T
log
always remains in the memory (ring memory principle).
The log contents are cleared if the length of the logging interval is changed.
Example
If 1 logging channel is used:
• T
log
= 1000 × 1 s = 1 000 s ≈15 min
• T
log
= 1000 × 10 s = 10 000 s ≈ 3 h
• T
log
= 1000 × 80 s = 80 000 s ≈ 1 d
• T
log
= 1000 × 3 600 s = 3 600 000 s ≈ 41 d
Clear logging
Navigation
Expert → Diagnostics → Data logging → Clear logging (0855)
Prerequisite
The
Extended HistoROM
application package is available.
The software options currently enabled are displayed in the
SW option overv.
Description
Use this function to clear the entire logging data.
Selection
• Cancel
• Clear data
Factory setting
Cancel
Additional information
Selection
• Cancel
The data is not cleared. All the data is retained.
• Clear data
The logging data is cleared. The logging process starts from the beginning.
Data logging
Navigation
Expert → Diagnostics → Data logging → Data logging (0860)
Description
Use this function to select the data logging method.
Selection
• Overwriting
• Not overwriting