Short Instructions
..
Quick configuration for level measurement
Function
Matrix
Action
1
Reset transmitter
V9H5
●
Enter 671
Press »E« to register entry
-
Omit if commissioned as in Section 4.1
2
»Empty« calibration*
V0H1
●
Fill vessel 0
…
40% full (probe covered)
Enter level in %, m, ft, etc.
Press »E« to register entry
3
»Full« calibration*
V0H2
●
Fill vessel 60
…
100% full
Enter level in %, m, ft, etc.
Press »E« to register entry
4
0/4 mA signal
V0H3
V0H5
V0H6
●
Enter 0 for 0
…
20 mA signal, 1 for 4
…
20 mA signal
Press »E« to register entry
●
Enter level for 0/4 mA signal (if not 0)
Press »E« to register entry
●
Enter level for 20 mA signal (if not 100)
Press »E« to register entry
5
Relays 1a and 1b
V1H0
V1H1
●
Enter level for switching in calibration units
Press »E« to register entry
●
Enter fail-safe mode: 0 = minimum, 1 = maximum
Press »E« to register
6
Relays 2a and 2b
V1H5
V1H6
V1H9
●
Enter level for switching in calibration units
Press »E« to register entry
●
Enter fail-safe mode: 0 = minimum, 1 = maximum
Press »E« to register
●
Enter 1 = Relay 2 assigned to channel 1
Press »E« to register entry
* Can be performed in reverse order
+
+
4 yellow relay LEDs,
1 red alarm LED,
1 green operation LED LC display
Operation
Selects vertical matrix position
Selects horizontal matrix position
Select position V0H0
Selects next digit
Move decimal point
Increases value of digit
Decreases value of digit
Registers entry
Short Instructions