Technical data
30
Hauser
NOTICE
Danger of drying out
►
Only store the sensor with the protection cap (filled with tap water).
10.3.2 Ambient temperature range
–5 to +135
˚
C (23 to 175
˚
F)
10.3.3 Ingress protection
IP 68 (10 m (33 ft) water column at 25
˚
C (77
˚
F) during 45 days, 1 mol/l KCl)
10.4 Process
10.4.1 Process temperature
–5 to +135
˚
C (23 to 175
˚
F)
10.4.2 Process pressure
ambient pressure to 12 bar (174 psi)
10.4.3 Temperature-pressure chart
10.4.4 Minimum flow
COS22-*1 (standard sensor): 0.02 m/s (0.07 ft/s)
COS22-*3 (trace ensor): 0.1 m/s (0.33 ft/s)
a0015193
Fig. 10: Temperature-pressure diagram