The following description applies for both sides, i.e. the detector side and the lamp side.

Always change O-rings or optical windows

 1)

 on both sides.

3.

Remove the 4 Allen screws (1/8" or 3 mm) from the window ring. Make sure to loosen

the screws evenly and alternately around the window ring.

4.

Remove the window ring .

5.

Gently push the optical window out of the assembly. If the window jams, apply some

acetone around the window seal (O-ring) and wait a few minutes for it to take effect.

This should help release the window. 

The seal cannot be reused afterwards!

Checking or replacing optical windows and seals

1. Check the window area for residue or fouling. Clean if necessary.
2. Check the optical windows for signs of chipping or abrasion.

 Replace the windows if signs of chipping/abrasion are found.

3. Dispose of all the O-rings and replace them with new O-rings from the relevant

maintenance kit.

1)

Optical windows only need to be replaced if they are damaged.

Содержание OUSAF46

Страница 1: ...Products Solutions Services Operating Instructions OUSAF46 Optical sensor with the OUA260 flow assembly for the measurement of UV absorption BA01349C 07 EN 03 18 71430357 2018 10 31...

Страница 2: ...ation conditions 9 5 2 Mounting the sensor 11 5 3 Post installation check 12 6 Electrical connection 12 6 1 Connecting the sensor 13 6 2 Lamp voltage 14 6 3 Ensuring the degree of protection 14 6 4 Po...

Страница 3: ...ituation Failure to avoid the dangerous situation can result in a fatal or serious injury LCAUTION Causes consequences If necessary Consequences of non compliance if applicable Corrective action This...

Страница 4: ...Operating Instructions provided must be carried out only directly at the manufacturer s site or by the service organization 2 2 Designated use The sensor measures the spectral absorption of process li...

Страница 5: ...damaged 3 Do not operate damaged products and protect them against unintentional operation 4 Label damaged products as defective During operation If faults cannot be rectified products must be taken o...

Страница 6: ...bsorption Extinction coefficient c Concentration OPL Optical path length A light source emits radiation through the medium and the incident radiation is measured on the detector side The subsequent co...

Страница 7: ...otected against impact and moisture The original packaging offers the best protection Make sure to comply with the permitted ambient conditions If you have any questions please contact your supplier o...

Страница 8: ...n option in the Product Configurator for the transmitter the complete measuring system transmitter sensor cable is factory calibrated and shipped as one package If you have any queries Please contact...

Страница 9: ...uring system comprises Sensor photometer OUSAF46 Transmitter e g Liquiline CM44P Cable set e g CUK80 Assembly OUA260 1 2 3 7 5 4 6 2 Example of a measuring system with a photometer sensor 1 pipe 5 Flo...

Страница 10: ...tandard lamp 146 1 5 75 Detector type Dimension B in mm inch Standard version with test filter 80 3 15 Easycal 94 3 70 The total length of the sensor module is derived from the lengths of the lamp the...

Страница 11: ...Forbidden mounting angle 5 2 Mounting the sensor The sensors have been specifically designed so that they can be installed in the process together with a flow assembly such as the OUA260 The flow ass...

Страница 12: ...y OUA260 the flow assembly is ready mounted on the sensor upon delivery The sensor is ready for use immediately If the sensor and assembly are ordered separately you must mount the sensor as follows 1...

Страница 13: ...set must be ordered separately BK GY RD A B YE YE BK BK GY RD BK B GY WH GN GN GY WH A0033423 5 OUSAF46 connecting cable A Light source lamp power supply B Signals of measurement and reference detect...

Страница 14: ...e mechanical and electrical connections which are described in these instructions and which are necessary for the required designated use may be carried out on the device delivered Exercise care when...

Страница 15: ...e nameplate Visual inspection Are the installed cables strain relieved and not twisted Has the cable been routed without loops and cross overs Check that it is firmly seated by pulling gently Are the...

Страница 16: ...born child and intensify fires Potentially life threatening if inhaled toxic if swallowed harmful if it comes into contact with skin Causes severe skin burns and severe eye damage When working with po...

Страница 17: ...and the actual absorbance at individual wavelengths is determined It is very important that you use the actual values of the optical Easycal filter These values are given in the calibration certificat...

Страница 18: ...played 2 Insert the reference rod into the detector housing Make a note of the measured value You can check the calibration any time later on by filling the flow assembly with water again and re inser...

Страница 19: ...ed if it is damaged Replace O rings in contact with the medium The replacement of O rings in contact with the medium depends on the specific requirements of the process Never re use a used O ring 8 2...

Страница 20: ...down 30 minutes 1 Turn the lamp module counterclockwise to remove it from the flow assembly 2 Use a 1 open ended wrench With the wrench hold the base plate of the cable connector in position and unscr...

Страница 21: ...lations for lamps containing mercury 5 Check whether the cables of the lamp module are worn or whether the reflector is damaged Inserting a new lamp When working with the lamp use a cloth for lenses o...

Страница 22: ...ng and tighten it by hand 6 Mount the complete lamp module back into the flow assembly After replacing the lamp the system needs to be adjusted In addition you must reset the operating hours counter f...

Страница 23: ...the exact same way as described for replacing the mercury lamp 20 1 2 3 4 9 Replacing the reference filter 1 Screw on circuit board 2 O ring 3 Reference filter 4 Lamp holder 1 Remove both screws item...

Страница 24: ...th a standard detector Please note that the measurement filters are not identical Their arrangement in the measuring channels must correspond to the arrangement of the reference filters in the lamp Me...

Страница 25: ...OUSAF46 Maintenance Endress Hauser 25 3 Remove housing along with lens assembly 4 Release two anodized black screws 5 Carefully pull apart Do not damage or break wires 6 Release screws on board...

Страница 26: ...s Hauser 7 Remove board 8 Loosen securing screw 9 Press out filter The wavelength of the filter is indicated on the socket 10 Use filters that have the same wavelength The arrow on the socket indicate...

Страница 27: ...place the measurement filter in the same way release screw press out filter insert new filter tighten screw 13 Re install board Pay attention to wires 14 Re secure the filter holder using the two anod...

Страница 28: ...mounted the sensor in a VARIVENT flow assembly using the CUA261 adapter refer to the Operating Instructions of the adapter for information on removal and on replacing the optical windows Removing opt...

Страница 29: ...assembly If the window jams apply some acetone around the window seal O ring and wait a few minutes for it to take effect This should help release the window The seal cannot be reused afterwards Check...

Страница 30: ...assembling and assembling the windows 9 Repair 9 1 Return The product must be returned if repairs or a factory calibration are required or if the wrong product was ordered or delivered As an ISO certi...

Страница 31: ...set Pre terminated and labeled cables for connecting analog photometer sensors Product Configurator on the product page www endress com cuk80 10 3 Calibration Kit OUSAF46 EasyCal retrofit kit Patented...

Страница 32: ...X 11 3 Process 11 3 1 Process temperature 0 to 90 C 32 to 194 F continuous Max 130 C 266 F for 2 hours 11 3 2 Process pressure Max 100 bar 1450 psi absolute depending on the material pipe size and pro...

Страница 33: ...e connector ends Nickel plated brass 11 4 4 Light source Low pressure mercury lamp Lamp operating life typically 3000 h The lamp will not operate at full capacity until a warm up period of 30 minutes...

Страница 34: ...Maintenance schedule 19 Measured variable 31 Measuring principle 6 Measuring range 31 Measuring system 9 Mounting angles 11 Mounting the sensor 11 N Nameplate 7 O Occupational safety 4 Operational saf...

Страница 35: ......

Страница 36: ...www addresses endress com 71430357 71430357...

Отзывы: