Maintenance
OUSAF46
22
Hauser
3.
Retighten the securing screw.
This screw is only used to hold the lamp in position. Therefore, do not tighten it too
tightly as to do so could damage or break the lamp.
4. Make sure that the lamp contact area is completely clean and dry before re-closing.
5.
Screw the lamp and reflector group back into its housing and tighten it by hand.
6.
Mount the complete lamp module back into the flow assembly.
After replacing the lamp, the system needs to be adjusted. In addition, you must reset the
operating hours counter for the lamp on the CM44P:
CAL
/
Photometer
/
Lamp operating
time
/
Reset
.
8.4
Replacing the reference filter
L
WARNING
Dangerous voltage
Risk of serious or fatal injury from electric shock!
‣
Before performing any maintenance work on the lamp, remove all the cables and
disconnect the power supply to the sensor.
‣
Only replace the filter in a current-free state.