background image

TI281T/02/en

Technical Information

Omnigrad S 

-

TR62

RTD thermometer EEx-d or EEx-ia certified, replaceable inset , complete 
of niple and union to thermowell connection
PCP (4...20 mA), HART® or PROFIBUS-PA® electronics

Range of uses

The Omnigrad S TR62 is an RTD industrial thermometer,
developed for heavy duty and/or generic industrial applica-
tions.
In compliance to EN 50014/18 (ATEX certification) it is there-
fore particularly suitable also for hazardous areas.
When required, it’s also available with a transmitter(PCP,
HART® or PROFIBUS-PA®) into the housing. 
The TR62 is available in several standard versions and different
configurations, can also be configured with specific dimensions
and characteristics depending on process requirements.
The installation on the plants, require separately order of the
thermowell (form pipe or from bar-stock).

Application areas

• Chemicals industry
• Energy industry
• Gas Processing industry
• Petrochemical industry
• General industrial services 

Features and benefits

• Customized immersion length
• Aluminium housing, with protection grade from IP66 / IP 68
• PCP, HART® and PROFIBUS-PA®, (4...20 mA 2-wire trans-

mitters)

• Thermoresistance insert insulated with mineral oxide cable 

(MgO cable) diameter: 3 or 6 mm

• Pt100 sensing element with accuracy in class A  or 1/3 DIN B 

(IEC 60751)

• The Pt100 available are: wire-wound WW (-200...600°C) or 

thin-film TF (-50...400°C)

• Single or double Pt100 to 2,3 or 4 wires connection
• ATEX II 1/2 GD EEx-ia IIC certification
• ATEX II 2 G EEx-d IIC certification

4 0

Содержание Omnigrad S TR62 Series

Страница 1: ...nfigured with specific dimensions and characteristics depending on process requirements The installation on the plants require separately order of the thermowell form pipe or from bar stock Applicatio...

Страница 2: ...S TR62 temperature sensor is based on the following standards EN 50014 18 assembly Neck ASME style nipple and 3 elements coupling IEC 60751 insert The housing is in painted aluminium alloy it is suita...

Страница 3: ...he probe see fig 2A 2C Insulation of the outer part of the sensor reduces the effect produced by a low immersion Alternatively it is also possible to adopt a tilted installation see fig 2B 2D For a be...

Страница 4: ...term stability System components Housing The protection housing our TA21H commonly referred to the connection head is used to contain and protect the terminal block or the transmitter and to join the...

Страница 5: ...the transmitters can be easily programmed using a PC In the case of PROFIBUS PA transmitters E H recommends the use of PROFIBUS dedicated connectors The Weidm ller type is provided as a standard opti...

Страница 6: ...smitter can be executed with 3 wires as well by avoiding to connect whichever of the terminals fig 5 The configuration Pt100 double with 2 wires is only available for the ATEX intrinsically safe certi...

Страница 7: ...rective is not requested Material certification The material certificate EN 10204 3 1 can be directly selected from the sale structure of the product and refers to the parts of the sensor in contact w...

Страница 8: ...T M D 104 mm SS 316 NU 1 2 NPT F E 156 mm SS 316 NUN 1 2 NPT M F 77 mm A 105 N 1 2 NPT M G 117 mm A 105 N 1 2 NPT M H 104 mm A 105 NU 1 2 NPT F J 156 mm A 105 NUN 1 2 NPT M Y Special version to be spe...

Страница 9: ...pan 0 250 C A21 TMT180 A21 fixed range accuracy 0 2 K span limit 200 650 C from to C A22 TMT180 A22 fixed range accuracy 0 1 K span limit 50 250 C from to C F11 TMT181 A PCP 2 wire isolated programmab...

Страница 10: ...e head transmitter iTEMP PROFIBUS PA TMT184 TI079R 09 en RTD insert for temperature sensors Omniset TPR100 TI268T 02 en RTD insert for temperature sensors Omniset TPR300 TI290T 02 en Safety instructio...