![Endress+Hauser Omnigrad M TR13 Скачать руководство пользователя страница 7](http://html1.mh-extra.com/html/endress-hauser/omnigrad-m-tr13/omnigrad-m-tr13_technical-information_2408484007.webp)
Omnigrad M TR13, TC13
7
Response time
Calculated at an ambient temperature of approx. 23 °C by immersing in running water (0.4 m/s flow
rate, 10 K excess temperature):
Complete assembly:
Thermometer
type
Diameter
t
(x)
Reduced tip
Tapered tip
Straight tip
Resistance
thermometer
(measuring probe
Pt100, TF/WW)
9 mm (0.35 in)
t
50
7.5 s
11 s
18 s
t
90
21 s
37 s
55 s
11 mm (0.43 in)
t
50
7.5 s
not available
18 s
t
90
21 s
not available
55 s
12 mm (0.47 in)
t
50
not available
11 s
38 s
t
90
not available
37 s
125 s
Thermo-
meter type
Diameter t
(x)
Grounded
Ungrounded
Reduced
tip
Tapered
tip
Straight
tip
Reduced
tip
Tapered
tip
Straight
tip
Thermo-
couple
9 mm
(0.35 in)
t
50
5.5 s
9 s
15 s
6 s
9.5 s
16 s
t
90
13 s
31 s
46 s
14 s
33 s
49 s
11 mm
(0.43 in)
t
50
5.5 s
not
available
15 s
6 s
not
available
16 s
t
90
13 s
not
available
46 s
14 s
not
available
49 s
12 mm
(0.47 in)
t
50
not
available
8.5 s
32 s
not
available
9 s
34 s
t
90
not
available
20 s
106 s
not
available
22 s
110 s
Response times for insert without transmitter.