Connection
•
Put the 4 free cores into the connector supplied so that after connecting
the colours correspond. The plug connector has cutting terminals so that
the cores must not be bared.
•
Crimp the plug connector together (e.g with universal pliers or a
multiwrench). See Fig. 23.
•
Cut off all remaining core ends.
•
Plug the connector on the end of the probe cable into the connector on
the end of the plug base. See Fig. 24.
•
Recheck: Do the core colours correspond with each other?
•
Plug the plug base for the electronic insert with mounting on the central
threaded rod in the housing.
The plug base should lie approximately opposite the ground screw.
•
Put the plug connector and cores right at the bottom of the housing.
•
Plug the electronic insert into the plug base and secure with the central
screw while making sure that the cable gland remains free.
•
Screw on housing cover.
Fig. 23
Left:
The four cores are inserted in the plug
connector from the side indicated.
Right:
The plug connectors are pressed together
and the core ends cut off.
Fig. 24
The two plug connectors are pressed
together.
Do the core colours correspond with each
other?
Nivocompact FTC 831
Shortening the Cable