Incoming acceptance and product identification
NAR300 system for high temperature
18
Hauser
Order Code
Seri. no.
Hauser Yamanashi Co., Ltd
Yamanashi 406-0846 Made in Japan
NP-2739
1
2
3
4
NRR261
XP-AIS Class I, Div. 1,2, Gp. C, D, T6
Class I, Zone 1[1], AEx db [ia] IIB T6
Ambient temperature: -20°C ~ + 60°C
Intrinsically safe circuit: NEMA 4X, IP67
Uo = 28 V Io = 85 mA Po = 595 mW Co = 0.083
F Lo = 2.4mH
Non intrinsically safe circuit:
Power supply:
Um : AC 250 V 50/60 Hz, DC 250 V
Contact output : 5 A 250 V AC, 5 A 30 V DC
Manufacturing date:
Conduit entry of the main body: M26 x 1.5
μ
Caution: A seal shall be installed within 18 inches of the enclosure.
: Do not modify internal parts or circuits.
: Use supply wires suitable 70°C minimum.
: Do not open the cover when energized.
: Refer to control drawing Ex1088-1282-*.
WARNING: AFTER DE-ENERGIZING, DELAY 10 MINUTES BEFORE OPENING.
A
Order Code
Seri. no.
Hauser Yamanashi Co., Ltd
Yamanashi 406-0846 Made in Japan
NP-2738-1
1
2
3
4
NRR261
ATEX: II 2G Ex db [ia Ga] IIB T6 Gb
FM 14ATEX0048X
IECEx: Ex db [ia] IIB T6 Gb
IECEx FMG 14.0024X
Ambient temperature: -20°C ~ + 60°C NEMA 4X, IP67
Intrinsically safe circuit:
Uo = 28 V Io = 85 mA Po = 595 mW Co = 0.083
F Lo = 2.4mH
Non intrinsically safe circuit:
Power supply:
Um : AC 250 V 50/60 Hz, DC 250 V
Contact output : 5 A 250 V AC, 5 A 30 V DC
Manufacturing date:
Conduit entry of the main body: M26 x 1.5
μ
Caution: Do not modify internal parts or circuits.
: Use supply wires suitable 70°C minimum.
: Do not open the cover when energized.
: Refer to control drawing Ex1088-1282-*.
WARNING: AFTER DE-ENERGIZING, DELAY 10 MINUTES BEFORE OPENING.
B
A0039867
7
Nameplates for NRR261
A
NRR261 nameplate for FM (NAR300 separate type)
B
NRR261 nameplate for ATEX/IECEx (NAR300 separate type)
1
Order code
2
Serial number
3
Power-supply voltage
4
Manufacturing date
Содержание NAR 300
Страница 50: ......
Страница 51: ......
Страница 52: ...www addresses endress com 71559868 71559868...