MMP20 SONO-DIS (option D)
Page 50
Hauser
2-point calibration:
In a 2-point calibration, two material samples of different moisture content are measured and a
linear equation (f(x)=mx+b) is then calculated from this information. While a higher-value
polynomial is useful to achieve better accuracy, the linear equation often suffices, particularly in
the lower moisture range, and delivers very good results.
NOTE:
You require two material samples with different moisture values to perform a 2-point
material calibration. The moisture values must be determined with another method
(kiln-drying, or similar) before the calibration. Make sure to adhere to the following
sequence: first "lower moisture value" (drier material) and then "upper moisture value"
(wetter material).
Procedure:
At the start of the calibration, the calibration memory
(01
–
15) to be overwritten must be selected with the
"Up"
and "Down"
keys. Press the "Measure" key
to accept the setting.
Then set the percentage moisture of the lower moisture
value using the "Up"
and "Down"
keys, and again
press the "Measure" key
to confirm your entries.
To start measuring the material, press the "Measure"
key
again. Four measurements are taken to increase
the accuracy. An average of these measurement values is
then calculated. The measuring time is approximately
20 seconds. When the measurement is finished, the
measured pulse transit time is displayed briefly.
NOTE
: Before you start measuring, make sure that the
rods of the probe are fully immersed in the
material to be measured. The probe must remain
in the material for the entire duration of the
measurement and should not be moved.