Micropilot S FMR 533
10 Technical data
Hauser
87
Auxiliary energy
Electrical connection
Housing T 12 with separate terminal compartment.
Load HART
Minimum load for HART communication: 250
Ω
Cable entry
Cable gland: M20x1.5 or Pg13.5
Cable entry: G ½ or ½ NPT
Supply voltage
Power consumption
Max. 330 mW at 16 V, max. 500 mW at 24 V, max. 600 mW at 30 V.
Current consumption
Max. 21 mA (50 mA inrush current).
Power supply
For stand alone operation recommended via e.g. E+H RN 221 N/Z.
Performance characteristics
Reference operating
conditions
According to OIML R85:
• Temperature = –25…+55 °C (-13…+131 °F)
• Atmospheric pressure
• Relative humidity (air) = 65 % ±15%
• Medium properties: e.g. medium with good reflectivity and calm surface.
• Tank diameter: signal beam hits the tank wall only at one side.
Note!
No major interference reflections inside the signal beam.
Maximum measured
error
Absolute accuracy: 0.5 mm (2
σ
value)
Inventory Control Versios: ±3 mm
Proof of accuracy for
custody transfer
versions
The accuracy of each Micropilot S is established through a calibration
certificate that records the absolute and relative error at 10 equidistant points
during the final test. A Laser Interferometer (Jenaer Messtechnik ZLM 500)
with an absolute accuracy of 0.1 mm is used as a reference for the free space
measurements with FMR 533.
Each Micropilot S is delivered with the PTB and NMi type approval. Additional
initial factory verifications for custody applications are available on demand
for all radar instruments FMR 53x.
Maximum fill speed
By the first pass trough of measuring range: 100 mm/min., thereafter
unlimited.
Non-repeatability
0,3 mm (1/64")
Hysteresis
0.3 mm
Resolution
digital: 0.1 mm
analogue: 0.03 % of measuring range
Settling time
Typical 15 sec
Long-term drift
The long-term drift is within the specified accuracy.
Influence of ambiente
temperature
Within the specified accuracy according to OIML R85.
Software reliability
The software used in the radar instruments FMR 53x fulfills the requirements
of OIML R85. This particularly includes:
• cyclical test of data consistency
• non-volatile memory
• segmented data storage
The radar instruments Micropilot S continuously monitor the compliance with
accuracy requirements for custody transfer measurements according to
OIML R85. If the accuracy cannot be maintained, a specific alarm is
generated on the local display and via the digital communication.