Operating menu
Micropilot NMR84
138
Hauser
"Analog I/O" submenu
There is a
Analog I/O
submenu for each Analog I/O module of the device. This
submenu refers to terminals 1 to 3 of this module (an analog input or output). For
terminals 4 to 8 (always an analog input) refer to → 132.
D
E
F
C
B
A
1
1
1
1
3
2
2
4
1
HR
CDI
WP
on
SIM
2
2
3
3
4
4
1
1
2
2
3
3
4
4
5
5
6
6
7
7
8
8
POWER
i
B
B1-3
B4-8
1 2 3
4 5 6 7 8
C
1 2 3
4 5 6 7 8
C1-3
C4-8
A0032464
47
Terminals for the "Analog I/O" submenu ("B1-3" or "C1-3", respectively)
Navigation
Setup → Advanced setup → Input/output → Analog I/O
Operating mode
Navigation
Setup → Advanced setup → Input/output → Analog I/O → Operating mode
Description
Defines the operating mode of the analog I/O module.
Selection
• Disabled
• 4..20mA input
• HART 4..20mA input
• HART master
• 4..20mA output
• HART slave +4..20mA output
Factory setting
Disabled
Additional information
Read access
Operator
Write access
Maintenance
Meaning of the options
Direction of signal
Type of signal
Disabled
-
-
4..20mA input
Input from 1 external device
Analog (4...20mA)
HART 4..20mA input
Input from 1 external device
• Analog (4...20mA)
• HART
HART master
Input from up to 6 external devices
HART
Содержание Micropilot NMR84
Страница 177: ...Micropilot NMR84 Operating menu Endress Hauser 177 Additional information Read access Operator Write access Maintenance ...
Страница 204: ...Operating menu Micropilot NMR84 204 Endress Hauser Additional information Read access Operator Write access Maintenance ...
Страница 219: ...Micropilot NMR84 Operating menu Endress Hauser 219 Additional information Read access Operator Write access Maintenance ...
Страница 238: ...Operating menu Micropilot NMR84 238 Endress Hauser Read access Operator Write access Maintenance ...
Страница 255: ......
Страница 256: ...www addresses endress com 71545626 71545626 ...