Micropilot NMR81
Diagnostyka i usuwanie usterek
Hauser
97
Numer
diagnostyczny
Krótki tekst
Działanie naprawcze
Sygnał
statusu
[z fabryki]
Reakcje
diagnostyczne
[z fabryki]
442
AIO 1 … 2 current output
warning
1. Check process
2. Check current output
settings
C
Warning
443
AIO 1 … 2 Input not HART
compatible
Change PV source or AIO
input source.
C
Warning
452
HyTD correction value
1. Check device
configuration.
2. Check wiring.
C
Warning
452
CTSh
C
Warning
452
HTG
C
Warning
452
HTMS
C
Warning
484
Failure mode simulation
Deactivate simulation
C
Alarm
495
Diagnostic event
simulation
Deactivate simulation
C
Warning
500
AIO C1-3 source no longer
valid
Change input source
C
Warning
501
Level source no longer
valid
Change input source
C
Warning
502
GP1 source no longer valid Change input source
C
Warning
503
GP2 source no longer valid Change input source
C
Warning
504
GP3 source no longer valid Change input source
C
Warning
505
GP4 source no longer valid Change input source
C
Warning
506
Water level source no
longer valid
Change input source
C
Warning
507
Liquid temp source no
longer valid
Change input source
C
Warning
508
Vapor temperatur source
no longer valid
Change input source
C
Warning
509
Air temperature source no
longer valid
Change input source
C
Warning
510
P1 source no longer valid
Change input source
C
Warning
511
P2 source no longer valid
Change input source
C
Warning
512
P3 source no longer valid
Change input source
C
Warning
513
Upper density source no
longer valid
Change input source
C
Warning
514
Middle density source no
longer valid
Change input source
C
Warning
515
Lower density source no
longer valid
Change input source
C
Warning
516
Gauge command source no
longer valid
Change input source
C
Warning
517
Gauge status source no
longer valid
Change input source
C
Warning
518
Average density source no
longer valid
Change input source
C
Warning
519
Upper interface source no
longer valid
Change input source
C
Warning
520
Lower interface source no
longer valid
Change input source
C
Warning
521
Bottom level source no
longer valid
Change input source
C
Warning