Micropilot FMR51, FMR52 FOUNDATION Fieldbus
Important document information
Hauser
11
If the device is prepared for the remote display FHX50 (product structure: feature 030:
Display, Operation", option L or M), the Ex marking of some certificates changes according
to the following table
:
Feature 010
("Approval")
Feature 030 ("Display,
Operation")
Ex marking
BG
L or M
ATEX II 3G Ex nA [ia Ga] IIC T6-T1 Gc
BH
L or M
ATEX II 3G Ex ic [ia Ga] IIC T6-T1 Gc
B3
L or M
ATEX II 1/2G Ex d [ia] IIC T6-T1 Ga/Gb,
ATEX II 1/2D Ex ta [ia Db] IIIC Txx°C Da/Db
IG
L or M
IECEx Ex nA [ia Ga] IIC T6-T1 Gc
IH
L or M
IECEx Ex ic [ia Ga] IIC T6-T1 Gc
I3
L or M
IECEx Ex d [ia] IIC T6-T1 Ga/Gb,
IECEx Ex ta [ia Db] IIIC Txx°C Da/Db
MH
L or M
Ex ic [ia Ga] IIC T6 Gc
NG
L or M
NEPSI Ex nA [ia Ga] IIC T6-T1 Gc
NH
L or M
NEPSI Ex ic [ia Ga] IIC T6-T1 Gc
N3
L or M
NEPSI Ex d [ia] IIC T6-T1 Ga/Gb, DIP A20/21 [ia D] TA, Txx°C
IP6X
1)
The marking of certificates not mentioned in this table are not affected by the FHX50.