Technical data
Memosens Wave CAS80E
40
Hauser
Measured
variable
2 mm (0.08 in) gap
10 mm (0.4 in) gap
50 mm (1.97 in) gap
SAC
1 1/m
0.2 1/m
0.04 1/m
CODeq
2 mg/l
0.4 mg/l
0.08 mg/l
TOCeq
1 mg/l
0.2 mg/l
0.04 mg/l
BODeq
0.5 mg/l
0.1 mg/l
0.02 mg/l
Nitrate NO3-N
1 mg/l
0.2 mg/l
0.04 mg/l
APHA Hazen true 75 Hazen
15 Hazen
3 Hazen
APHA Hazen
apparent
32.5 Hazen
7.5 Hazen
1.5 Hazen
1)
A minimum path length of 25 mm (0.98 in) is required in US Standard Methods 2120C (Single
Wavelength Method) 23. Edition
Drinking water
Measured
variable
2 mm (0.08 in) gap
10 mm (0.4 in) gap
50 mm (1.97 in) gap
Turbidity
12.5 FAU
2.5 FAU
0.5 FAU
TSS
12.5 mg/l
2.5 mg/l
0.5 mg/l
SAC
1 1/m
0.2 1/m
0.04 1/m
TOCeq
1 mg/l
0.2 mg/l
0.04 mg/l
Nitrate NO3-N
1 mg/l
0.2 mg/l
0.04 mg/l
APHA Hazen true 75 Hazen
7.5 Hazen
3 Hazen
APHA Hazen
apparent
32.5 Hazen
15 Hazen
1.5 Hazen
1)
A minimum path length of 25 mm (0.98 in) is required in US Standard Methods 2120C (Single
Wavelength Method) 23. Edition
Surface water
Measured
variable
2 mm (0.08 in) gap
10 mm (0.4 in) gap
50 mm (1.97 in) gap
Turbidity
12.5 FAU
2.5 FAU
0.5 FAU
TSS
12.5 mg/l
2.5 mg/l
0.5 mg/l
SAC
1 1/m
0.2 1/m
0.04 1/m
CODeq
2 mg/l
0.4 mg/l
0.08 mg/l
BODeq
0.5 mg/l
0.1 mg/l
0.02 mg/l
Nitrate NO3-N
1 mg/l
0.2 mg/l
0.04 mg/l
Determination limit
The determination limits were determined for the individual measured variables in
ultrapure water under laboratory conditions based on DIN ISO 15839.
Wastewater treatment plant inlet
Measured
variable
2 mm (0.08 in) gap
10 mm (0.4 in) gap
50 mm (1.97 in) gap
TSS
66.7 mg/l
13.3 mg/l
2.7 mg/l
SAC
3.5 1/m
0.7 1/m
0.15 1/m