Repair
Memosens COS81E
32
Hauser
3. Reset the cap counters.
Warnings can be configured for the spot cap counters to assist with sensor
maintenance.
4.
A0043011
Screw on the new spot cap.
5. Calibrate the sensor or check the measuring function.→ 21
10.4 Disposal
The device contains electronic components. The product must be disposed of as electronic
waste.
‣
Observe the local regulations.
10.5 Checking the measurement function
1. Remove the sensor from the medium.
2. Clean and dry the spot cap.
3. After about 10 minutes, measure the oxygen saturation index in air (without
recalibration).
The measured value should be at 100 ± 2 % SAT.