Technical data
Memosens CLL47E
12
Hauser
9.2.2
Maximum measured error
Conductivity
In the range 5 µS/cm to 1 mS/cm
≤ 2 % of reading
In the range 1 mS/cm to 200 mS/cm ≤ 4 % of reading
Temperature
≤ 1.0 K, in measuring range 0 to 100 °C
(32 to 212 °F)
9.2.3
Repeatability
Conductivity
≤ 0.5 % of reading, in specified measuring range
Temperature
≤ 0.5 K
9.3
Environment
9.3.1
Ambient temperature range
–20 to 60 °C (–4 to 140 °F)
9.3.2
Storage temperature
-25 to +80 °C (-13 to +176 °F)
9.3.3
Conditions for outdoor use
If the sensor is used outdoors, the following conditions apply to maintain the confirmed
specification:
• Connection via CYK10-A052 cable, use of spacer (protection against loss)
• Maximum 30 minutes
• Maximum twice per week
• Maximum insertion depth 5 m (16.4 ft)
• Maximum medium temperature 50 °C (122 °F)
9.3.4
Humidity
5 to 95 %
9.3.5
Degree of protection
IP 68 / NEMA type 6P (1.9 m water column, 20 °C, 24 h)