Parameter → System → Minimum µC-Temperature
Description
This parameter is used as the minimum peak indicator and makes it possible to call up
retroactively the lowest electronics temperature measured.
Maximum µC-Temperature
Navigation
Parameter → System → Maximum µC-Temperature
Description
This parameter is used as the maximum peak indicator and makes it possible to call up
retroactively the highest electronics temperature measured.
Reset µC-Temperature
Navigation
Parameter → System → Reset µC-Temperature
Description
Use this function to set the maximum/minimum peak indicators for the µC temperature to
the temperature currently present. The minimum and maximum indicators have the same
value once the function has been executed.
Standard Command
Navigation
Parameter → System → Standard Command
Description
L
WARNING
If the user confirms the "Standard Command" function by pressing the "Reset to
factory settings" button, this causes the device to be immediately reset to the as-
delivered state.
Downstream processes can be affected. The behavior of the current outputs can change.
‣
Make sure that downstream processes are not started unintentionally.
To perform a reset there is no additional locking, e.g. device unlocking is not required to
perform a reset. Furthermore, the device status is also reset. Any customer-specific
configuration carried out at the factory is not affected by a reset (customer-specific
configuration remains).
The following parameters are
not
reset when a reset is performed:
• Minimum µC-Temperature
• Maximum μC-Temperature
• Last Diagnostic (LST)
• Operating hours
Device Access Locks.Data Storage Lock
Activation/deactivation of data storage
1)
The "Device Access Locks.Data Storage Lock" parameter is an IO-Link standard parameter. The name of the parameter may be available in the
language configured in the IO-Link operating tool that is used. The display depends on the particular operating tool.
Navigation
Parameter → System → Device Access Locks.Data Storage Lock