Select function group “setpoint 2” (SP 2).
Select the “Set Limit” function.
Change the setting, e.g. from 200.0 FNU to 120.0 FNU.
Confirm the entry.
The instrument advances to the next function.
Change the factory setting of
“Hi” (upper limit = Max Function) to
“Lo” (lower limit = Min Function).
Confirm the setting.
The instrument advances to the next function
(HYS = Hysteresis Setting).
Operation
liquisys CUM 252
22
Hauser