Liquiphant S FTL70, FTL71
14
Electronic insert FEL50A (PROFIBUS PA)
Power supply
Bus voltage: 9 to 32 V DC
Bus current:
• 12.5 mA +/– 1.0 mA (software version: 01.03.00, hardware version: 02.00)
• 10.5 mA +/– 1.0 mA (software version: 01.03.00, hardware version: 01.00)
Electrical connection
Two-wire connection for power supply and data transfer
L00-FTL5xxxx-04-05-xx-en-006
For connecting to PROFIBUS PA
Additional functions:
– Digital communication enables the
representation, reading and editing
of the following parameters:
Fork frequency, switch-on frequency,
switch-off frequency, switch-on time and
switch-off time, status, measured value, density
switch.
– Matrix locking possible
– Switch to WHG mode possible
(WHG approval).
– For a detailed description, see BA198F
– You can also visit
www.profibus.com for more information
L00-FTL5xxxx-04-05-xx-en-005
PA–
PA+
PROFIBUS PA
–
–
–
–
EX
EX
U– 9…32 V DC
...
1 2
FEL50A
Segment coupler
PLC
PROFIBUS DP
PROFIBUS PA
E
X
EX
PC with Commuwin II
Proficard or Profiboard