Liquiline System CA80HA
Operation
Hauser
99
Menu/Setup/Outputs/Binary output x:y
Function
Options
Info
Binary output
Selection
On
Signal type
Selection
Static signal
Function
Selection
Analyzer
Assignments
Function
=
Analyzer
Selection
Sample required SP1
Here select which binary outputs output the
system status for a measurement currently
running.
In the case of two-channel devices,
Sample
required SP2
is also displayed
1)
x:y = slot no. : input number
5. The output signal level
High
indicates that sample is required.
10.6.2 Binary input configuration
Menu/Setup/Inputs/Binary input x:y
Function
Options
Info
Binary input
Selection
• Off
• On
Factory setting
On
Switches the input on/off
Signal type
Selection
• Static signal
• PFM
Factory setting
Static signal
‣
Select the signal type.
Static signal
Use this setting to read out the position of, for
example, an on/off switch, an inductive
proximity switch or a PLC binary output.
Signal application: for measuring range
switching, acceptance of an external hold, as a
cleaning trigger or for controller activation
PFM
The PFM setting produces a pulse-frequency-
modulated signal that is subsequently available
on the device as a quasi-continuous process
value.
Example: Measuring signal of a flowmeter