Liquiline System CA80HA
Operation
Hauser
141
To calculate a difference, you must use two measured values with the same engineering unit.
Menu/Setup/Additional functions/Mathematical functions/MF1 to 8/Mode = Difference
Function
Options
Info
Calculation
Selection
• Off
• On
Factory setting
Off
On/off switch for the function
Y1
The options depend on the
sensors connected
Select the sensors and measured variables that
should function as the minuend (
Y1
) or
subtrahend (
Y2
).
Measured value
Y2
Measured value
Difference value
Read only
View this value in a user-defined measuring
screen or output the value via the current
output.
Mathematical function
assignment view
Overview of the configured functions
Redundancy
Use this function to monitor two or three sensors with redundant measurements. The
arithmetic average of the two closest measured values is calculated and output as the
redundancy value.
Menu/Setup/Additional functions/Mathematical functions/MF1 to 8/Mode = Redundancy
Function
Options
Info
Calculation
Selection
• Off
• On
Factory setting
Off
On/off switch for the function
Y1
The options depend on the
sensors connected
You can select a maximum of 3 different types
of sensor that output the same measured value.
Example for temperature redundancy
You have a pH sensor and an oxygen sensor at
inputs 1 and 2. Select the pH sensor as
Y1
and
the oxygen sensor as
Y2
.
Measured value
:
Select
Temperature
in each case.
Measured value
Y2
Measured value
Y3 (optional)
Measured value