Liquiline System CA80CR
Technical data
Hauser
181
15.7.10 Cleaning interval
1 to 90 days, depending on the application and ambient conditions
15.7.11 Maintenance interval
Every 3 - 6 months, depending on the application
15.7.12 Servicing requirements
• Weekly: visual inspection
• Quarterly: 1 hour
15.8 Environment
15.8.1 Ambient temperature range
5 to 40 °C (41 to 104 °F)
15.8.2 Storage temperature
-20 to 60 °C (-4 to 140 °F)
15.8.3 Height
Environment
≤
2000 m (< 6562 ft) above MSL
15.8.4 Humidity
10 to 95%, not condensing
15.8.5 Degree of protection
IP55 (cabinet, analyzer stand)
15.8.6 Electromagnetic compatibility
Interference emission and interference immunity as per EN 61326-1: 2013, class A for industry
15.8.7 Electrical safety
The device is suitable for connection to power supply networks in overvoltage category II.
The device meets technical safety requirements as per
• EN 61010-1 (3
rd
edition)
• UL Std. No. 61010-1 (3
rd
edition)
• CAN/CSA-C22.2 No. 61010-1-12
15.8.8 Pollution degree
• Cabinet version/analyzer stand: pollution degree 4
• Open installation: pollution degree 2