Operation
Liquiline System CA80SI
90
Hauser
Outputting the status of a cleaning function
Function = Cleaning(for sensors)
Function
Options
Info
Assignments
Selection
• None
• Depends on the type of
cleaning
Factory setting
None
Here you can specify how a cleaning function
should be displayed for the relay.
You have the following options depending on the
cleaning program that is selected (
Menu
/
Setup
/
Additional functions
/
Cleaning
:
•
Cleaning type
=
Standard clean
Cleaning 1 - Water
,
Cleaning 2 - Water
,
Cleaning 3 - Water
,
Cleaning 4 - Water
•
Cleaning type
=
Chemoclean
Cleaning 1 - Water
,
Cleaning 1 - Cleaner
,
Cleaning 2 - Water
,
Cleaning 2 - Cleaner
,
Cleaning 3 - Water
,
Cleaning 3 - Cleaner
,
Cleaning 4 - Water
,
Cleaning 4 - Cleaner
•
Cleaning type
=
Chemoclean Plus
4x
Cleaning 1 - %0V
, 4x
Cleaning 2 - %0V
Hold behavior
Selection
• Freeze last value
• Fixed value
• Ignore
Factory setting
Ignore
Freeze last value
The device freezes the last measured value.
Fixed value
You define a fixed measured value that is output
at the output.
Ignore
A hold has no effect.
1)
%0V is variable text which you can assign in
Menu
/
Setup
/
Additional functions
/
Cleaning
/
Chemoclean
Plus
/
Output label 1 ... 4
.
Outputting the "Measurement active" system status and the information "Sample
required"
Function = Analyzer
Function
Options
Info
Signal type
=
Static signal
Assignments
Function
=
Analyzer
Selection
• None
• Sample required SP1
• Measurement active SP1
Factory setting
None
If this option is selected, information is output as
to whether a measurement is active on the
selected measuring channel or an action is
started that requires sample (measurement,
calibration or cleaning).
10.6.3 PROFIBUS DP and PROFINET
Device variables (device → PROFIBUS/PROFINET)
Here you can specify which process values should be mapped to the PROFIBUS function
blocks and are thereby available for transmission via PROFIBUS communication.
You can define a maximum of 16 device variables (AI Blocks).
1. Define the data source.
You can choose from sensor inputs, current inputs and mathematical functions.
2. Choose the measured value to be transmitted.
3. Define how the device should behave in the hold state (for sensors). (Configuration
options of
Source of data
,
Measured value
and
Hold behavior