Commissioning

Levelflex FMP55 PROFIBUS PA

26

Hauser

6.

Setup  →  Tube diameter

 (only for "Tank type" = "Bypass / pipe")

 Specify the diameter of the bypass or stilling well.

7.

Setup  →  DC value

 Specify the dielectric constant of the upper medium

8.

Setup  →  Empty calibration

 Specify the empty distance E (distance from the reference point R to the 0% mark).

9.

Setup  →  Full calibration

 Specify the full distance F (distance from the 0% mark to the 100% mark).

10.

Setup  →  Level

 Displays the measured level L

L

.

11.

Setup  →  Interface

 Displays the interface height L

I

.

12.

Setup  →  Distance

 Displays the distance D

L

 between the reference point R and the level L

L

.

13.

Setup  →  Interface distance

 Displays the distance D

I

 between the reference point R and the interface L

I

.

14.

Setup  →  Signal quality

 Displays the signal quality of the analyzed level echo.

15.

Setup  →  Mapping  →  Confirm distance

 Make sure the vessel is completely empty. Then select 

Tank empty

 option.

NOTICE

Wrong measurement due to incorrect dielectric constant of the lower medium

If, for Operating mode = 

Interface with capacitance

, the lower medium is not water, the

dielectric constant (DC value) of the medium must be specified: Setup → Advanced setup → 

Interface  →  DC value lower medium

NOTICE

Wrong measurement due to incorrect empty capacitance

In the case of rod and rope probes in the bypass and if 

Operating mode

 = 

Interface with

capacitance

, a correct measurement is only possible once the empty capacitance has been

determined. For this purpose, select 

Confirm distance

 = 

Tank empty

 after installing the

probe when the tank is completely empty (Step 13 in the table above).

The empty capacitance is always calibrated ex works in the case of coax probes.

Содержание Levelflex FMP55 PROFIBUS PA

Страница 1: ...ey are not a substitute for the Operating Instructions pertaining to the device Detailed information about the device can be found in the Operating Instructions and the other documentation Available f...

Страница 2: ...cd Ser no www endress com deviceviewer Endress Hauser Operations App XXXXXXXXXXXX XXXXX XXXXXX XXX XXXX XX Serial number 1 3 2 A0023555 2 About this document 2 1 Symbols 2 1 1 Safety symbols DANGER Th...

Страница 3: ...hat must be connected to ground prior to establishing any other connections The ground terminals are located on the inside and outside of the device Inner ground terminal protective earth is connected...

Страница 4: ...owner operator Are familiar with federal national regulations Must have read and understood the instructions in the manual and supplementary documentation Follow instructions and comply with condition...

Страница 5: ...ule may rise up to 80 C 176 F When in operation the sensor may reach a temperature close to the medium temperature Danger of burns from contact with surfaces In the event of high medium temperatures e...

Страница 6: ...ted on the nameplate is no longer valid This may also impair the safe operation of the device 3 5 1 CE mark The measuring system meets the legal requirements of the applicable EU directives These are...

Страница 7: ...of the information on the measuring device and on the scope of the technical documentation pertaining to the device is displayed 4 3 Storage and transport 4 3 1 Storage temperature Permitted storage t...

Страница 8: ...connected to the electronics housing in the FMP55 with coax probe It is fixed onto the mounting flange with two cable ties during transportation These cable ties must not be released during transporta...

Страница 9: ...Levelflex FMP55 PROFIBUS PA Incoming acceptance and product identification Endress Hauser 9 1 2 4 3 A0015471...

Страница 10: ...in the bypass stilling well Coax probes can be mounted at any distance from the wall When mounting outdoors a weather protection cover can be used to protect the device against extreme weather condit...

Страница 11: ...he device 5 2 1 Mounting devices with a flange If a seal is used to mount the device use uncoated metal screws to ensure good electrical contact between the process flange and the probe flange 5 2 2 M...

Страница 12: ...d swinging of the weight as this could damage internal fittings in the vessel 5 2 3 Turning the transmitter housing To provide easier access to the connection compartment or display module the transmi...

Страница 13: ...rclockwise 2 Unscrew the electronics compartment cover and check the cover seal replace it if necessary Turning the display module 1 2 A0036401 1 Pull out the display module with a gentle rotational m...

Страница 14: ...5 Nm 3 mm A0021451 1 Screw down the cover of the electronics compartment 2 Turn the securing clamp 90 in the clockwise direction and using an Allen key 3 mm tighten the screw of the securing clamp on...

Страница 15: ...without integrated overvoltage protection 3 Connection switch output open collector terminals 3 and 4 with integrated overvoltage protection 4 Connection PROFIBUS PA FOUNDATION Fieldbus terminals 1 a...

Страница 16: ...wire FOUNDATION Fieldbus switch output G 2 wire PROFIBUS PA switch output Non hazardous Ex nA Ex nA ia Ex ic Ex ic ia Ex d ia XP Ex ta DIP CSA GP 9 to 32 V 3 Ex ia IS Ex ia Ex d ia IS XP 9 to 30 V 3...

Страница 17: ...ect the potential matching line to the outer ground terminal before applying the power supply Required tools accessories For devices with a cover lock Allen key AF3 Wire stripper When using stranded c...

Страница 18: ...shielded cables Connect the cable shield to the ground terminal 6 2 3 Plug in spring force terminals The electrical connection of device versions without an integrated overvoltage protection is via p...

Страница 19: ...partment 2 Turn the securing clamp 90 in the clockwise direction and using an Allen key 3 mm tighten the screw of the securing clamp on the connection compartment cover with 2 5 Nm 7 Operation options...

Страница 20: ...ymbol if an error is active 1 2 Measured value symbols 1 3 Measured value 1 4 Unit 2 Measured value display bar graph 1 value 2 1 Bar graph for measured value 1 2 2 Measured value 1 including unit 2 3...

Страница 21: ...isplay module 1 Operating keys Key assignment Key Navigate down in the picklist Edit the numerical values or characters within a function Key Navigate up in the picklist Edit the numerical values or c...

Страница 22: ...nt should generally be completely configured Setup Mapping Mapping of interference echos Setup Advanced setup Contains additional submenus and parameters For more customized configuration of the measu...

Страница 23: ...l display appears Calling up the menu via the context menu 1 Open the context menu 2 Press to navigate to the desired menu 3 Press to confirm the selection The selected menu opens 8 Commissioning 8 1...

Страница 24: ...eration Setup Main menu 0104 1 Display language English Espa ol Fran ais Display language English Deutsch 0104 1 Espa ol Fran ais Display language English Deutsch 0104 1 Betrieb Setup Hauptmen Sprache...

Страница 25: ...DI Interface distance distance from flange to DK2 E Empty calibration zero point LI Interface F Full calibration span DL Distance LN probe length LL Level UP Measured thickness upper layer 1 Setup Ta...

Страница 26: ...14 Setup Signal quality Displays the signal quality of the analyzed level echo 15 Setup Mapping Confirm distance Make sure the vessel is completely empty Then select Tank empty option NOTICE Wrong mea...

Страница 27: ......

Страница 28: ...www addresses endress com 71565964 71565964...

Отзывы: