Electrical connection
Levelflex FMP55 PROFIBUS PA
14
Hauser
Closing the cover of the electronics compartment
1.
2.
2.5 Nm
3 mm
A0021451
1. Screw down the cover of the electronics compartment.
2. Turn the securing clamp 90 ° in the clockwise direction and, using an Allen key (3 mm),
tighten the screw of the securing clamp on the electronics compartment cover with
2.5 Nm.
6
Electrical connection
6.1
Connecting requirements
6.1.1
Terminal assignment