Brief Operating Instructions

Levelflex FMP50

HART

Guided wave radar

These Brief Operating Instructions are not a substitute for the

Operating Instructions pertaining to the device.
Detailed information about the device can be found in the

Operating Instructions and the additional documentation.
Available for all device versions via

• Internet: 

www.endress.com/deviceviewer

• Smartphone/tablet: 

Hauser Operations app

KA01053F/00/EN/20.22-00

71567028

2022-04-05

Содержание Levelflex FMP50

Страница 1: ...titute for the Operating Instructions pertaining to the device Detailed information about the device can be found in the Operating Instructions and the additional documentation Available for all devic...

Страница 2: ...er no www endress com deviceviewer Endress Hauser Operations App XXXXXXXXXXXX XXXXX XXXXXX XXX XXXX XX Serial number 1 3 2 A0023555 2 About this document 2 1 Symbols 2 1 1 Safety symbols DANGER This s...

Страница 3: ...must be connected to ground prior to establishing any other connections The ground terminals are located on the inside and outside of the device Inner ground terminal protective earth is connected to...

Страница 4: ...ed by the plant owner operator Are familiar with federal national regulations Must have read and understood the instructions in the manual and supplementary documentation Follow instructions and compl...

Страница 5: ...may rise up to 80 C 176 F When in operation the sensor may reach a temperature close to the medium temperature Danger of burns from contact with surfaces In the event of high medium temperatures ensu...

Страница 6: ...on the nameplate is no longer valid This may also impair the safe operation of the device 3 5 1 CE mark The measuring system meets the legal requirements of the applicable EU directives These are lis...

Страница 7: ...nformation on the measuring device and on the scope of the technical documentation pertaining to the device is displayed 4 3 Storage and transport 4 3 1 Storage temperature Permitted storage temperatu...

Страница 8: ...Mounting Levelflex FMP50 HART 8 Endress Hauser 5 Mounting 5 1 Mounting requirements 5 1 1 Suitable mounting position A C 1 2 3 4 B A0012606 1 Installation conditions for Levelflex...

Страница 9: ...istance B between rod probes and internal fittings 3 300 mm 12 in When using more than one Levelflex Minimum distance between the sensor axes 100 mm 3 94 in Distance C from the end of the probe to the...

Страница 10: ...ly grounded or reliably insulated Use an insulated fastening kit if it is not otherwise possible to secure the probe with a reliably insulated connection Securing rod probes In the case of WHG approva...

Страница 11: ...ing devices with a threaded connection A0012528 Screw the device with the threaded connection into a sleeve or flange and then secure it to the process vessel via the sleeve flange When screwing into...

Страница 12: ...l Uncoil the rope slowly and lower it carefully into the vessel Make sure the rope does not bend or buckle Avoid uncontrolled swinging of the weight as this could damage internal fittings in the vesse...

Страница 13: ...g an Allen key 3 mm and turn the clamp 90 counterclockwise 2 Unscrew the electronics compartment cover and check the cover seal replace it if necessary Turning the display module 1 2 A0036401 1 Pull o...

Страница 14: ...s compartment cover with 2 5 Nm 6 Electrical connection 6 1 Connecting requirements 6 1 1 Terminal assignment Terminal assignment 2 wire 4 to 20 mA HART 1 2 A 1 2 B 1 2 3 3 A0036498 2 Terminal assignm...

Страница 15: ...uring device Terminal assignment 2 wire 4 to 20 mA HART switch output A 1 2 3 4 B 1 2 3 4 1 3 5 5 2 4 A0036500 4 Terminal assignment 2 wire 4 to 20 mA HART switch output A Without integrated overvolta...

Страница 16: ...e 7 Switch output open collector Terminal assignment 2 wire 4 to 20 mA HART 4 to 20 mA A 1 2 3 4 B 1 2 3 4 1 3 5 5 2 4 A0036500 6 Terminal assignment 2 wire 4 to 20 mA HART 4 to 20 mA A Without integr...

Страница 17: ...A195 or FieldXpert SFX350 SFX370 via VIATOR Bluetooth modem 4 Analog display unit observe maximum load 5 Cable screen observe cable specification 6 Measuring device 7 Analog display unit observe maxim...

Страница 18: ...ubox FXA195 or FieldXpert SFX350 SFX370 via VIATOR Bluetooth modem 4 Analog display unit observe maximum load 5 Cable screen observe cable specification 6 Measuring device 7 Supply voltage observe ter...

Страница 19: ...e or threaded connection or to the external ground terminal An easily accessible power switch must be installed in the proximity of the device The power switch must be marked as a disconnector for the...

Страница 20: ...Levelflex FMP50 HART 20 Endress Hauser 2 1 3 4 A0011175 12 Pin assignment of M12 plug 1 Signal 2 Not assigned 3 Signal 4 Ground 4 2 3 1 A0011176 13 Pin assignment of 7 8 plug 1 Signal 2 Signal 3 Not...

Страница 21: ...A 2 Feature 010 in the product structure 3 At ambient temperatures Ta 30 C a terminal voltage U 14 V is required to start the device with the minimum failure current 3 6 mA At ambient temperatures Ta...

Страница 22: ...ic Ex ic ia Ex d ia XP Ex ta DIP CSA GP 13 5 to 35 V 3 4 R V U0 10 13 5 24 5 20 30 0 500 35 A0034971 Ex ia IS Ex ia Ex d ia IS XP 13 5 to 30 V 3 4 1 Feature 020 of the product structure option B 2 Fe...

Страница 23: ...Feature 020 of the product structure option C 2 Feature 010 in the product structure 3 At ambient temperatures Ta 30 C a terminal voltage U 16 V is required to start the device with the minimum failur...

Страница 24: ...Instructions XA Use specified cable glands only Check to ensure that the power supply matches the information on the nameplate Switch off the power supply before connecting the device Connect the pot...

Страница 25: ...cables also fit ferrules 4 Firmly tighten the cable glands 5 Connect the cable according to the terminal assignment 2 1 4 3 2 1 4 3 A0034682 6 If using shielded cables Connect the cable shield to the...

Страница 26: ...osing the cover of the connection compartment 1 2 2 5 Nm 3 mm A0021491 1 Screw down the cover of the connection compartment 2 Turn the securing clamp 90 in the clockwise direction and using an Allen k...

Страница 27: ...ol if an error is active 1 2 Measured value symbols 1 3 Measured value 1 4 Unit 2 Measured value display bar graph 1 value 2 1 Bar graph for measured value 1 2 2 Measured value 1 including unit 2 3 Me...

Страница 28: ...lay module 1 Operating keys Key assignment Key Navigate down in the picklist Edit the numerical values or characters within a function Key Navigate up in the picklist Edit the numerical values or char...

Страница 29: ...should generally be completely configured Setup Mapping Mapping of interference echos Setup Advanced setup Contains additional submenus and parameters For more customized configuration of the measurem...

Страница 30: ...l display appears Calling up the menu via the context menu 1 Open the context menu 2 Press to navigate to the desired menu 3 Press to confirm the selection The selected menu opens 8 Commissioning 8 1...

Страница 31: ...ion Setup Main menu 0104 1 Display language English Espa ol Fran ais Display language English Deutsch 0104 1 Espa ol Fran ais Display language English Deutsch 0104 1 Betrieb Setup Hauptmen Sprache Deu...

Страница 32: ...ance E Empty calibration zero point L Level F Full calibration span 1 Setup Device tag Enter device tag 2 Setup Distance unit Select the length unit 3 Setup Tank type Select tank type 4 Setup Tube dia...

Страница 33: ...level L for verification purposes 9 Setup Distance Displays the distance D between the reference point R and the level L for verification purposes 10 Setup Signal quality Displays the signal quality o...

Страница 34: ......

Страница 35: ......

Страница 36: ...www addresses endress com 71567028 71567028...

Отзывы: