Proline Promass 80
Troubleshooting
Hauser
57
9.2
System error messages
Serious system errors are
always
recognized by the instrument as a "Fault message" and are shown
as a lightning flash (
$
) on the display! Fault messages immediately affect the inputs and outputs.
"
Caution!
In the event of a serious fault, a flowmeter might have to be returned to the manufacturer
for repair. Important procedures must be carried out before you return a flowmeter to
Hauser
Always enclose a duly completed "Declaration of contamination" form.
You will find a preprinted blank of this form at the back of this manual.
!
Note!
• The error types listed in the following correspond to the factory settings.
• See the information on
No.
Error message / Type
Cause
Remedy (spare part
63)
S = System error
$
= Fault message (with an effect on the outputs)
! = Notice message (without an effect on the outputs)
No. # 0xx
Hardware error
001
S: CRITICAL FAILURE
$
: # 001
Serious device error
Replace the amplifier board.
011
S: AMP HW EEPROM
$
: # 011
Amplifier:
Defective EEPROM
Replace the amplifier board.
012
S: AMP SW EEPROM
$
: # 012
Measuring amplifier: Error when accessing data of the
EEPROM
The EEPROM data blocks in which an error has occurred are
displayed in the TROUBLESHOOTING function.
Press Enter to acknowledge the errors in question; default values
are automatically inserted instead of the erroneous parameter
values.
!
Note!
The measuring device has to be restarted if an error has occurred
in a totalizer block (see error No. 111/CHECKSUM TOTAL.).
031
S: SENSOR HW DAT
$
: # 031
1.
S-DAT is not plugged into the amplifier board
correctly (or is missing).
2.
S-DAT is defective.
1.
Check whether the S-DAT is correctly plugged into the
amplifier board.
2.
Replace the S-DAT if it is defective.
Check that the new, replacement DAT is compatible with the
measuring electronics.
Check the:
- Spare part set number
- Hardware revision code
3.
Replace measuring electronics boards if necessary.
4.
Plug the S-DAT into the amplifier board.
032
S: SENSOR SW DAT
$
: # 032
No. # 1xx
Software error
121
A / C COMPATIB.
!: # 121
Due to different software versions, I/O board and
amplifier board are only partially compatible (possibly
restricted functionality).
!
Note!
– This message is only listed in the error history.
– Nothing is displayed on the display.
Module with lower software version has either to be actualized by
FieldCare with the required software version or the module has to
be replaced.
No. # 2xx
Error in DAT / no communication
251
COMMUNICATION I/O
$
: # 251
Internal communication fault on the amplifier board.
Remove the amplifier board.
261
COMMUNICATION I/O
$
: # 261
No data reception between amplifier and I/O board or
faulty internal data transfer.
Check the BUS contacts
Содержание HART Proline Promass 80