Flowmeter Proline 500 – digital
Electrical connection
Hauser
15
5.1.3
Connecting cable
Standard cable
A standard cable can be used as the connecting cable.
Standard cable
4 cores (2 pairs); pair-stranded with common shield
Shielding
Tin-plated copper-braid, optical cover ≥ 85 %
Cable length
Maximum 300 m (1
000 ft), see the following table.
Cross-section
Cable lengths for use in
Non-hazardous area,
Ex Zone 2, Class I, Division 2
Hazardous area,
Ex Zone 1, Class I, Division 1
0.34 mm
2
(AWG 22)
80 m (270 ft)
50 m (165 ft)
0.50 mm
2
(AWG 20)
120 m (400 ft)
60 m (200 ft)
0.75 mm
2
(AWG 18)
180 m (600 ft)
90 m (300 ft)
1.00 mm
2
(AWG 17)
240 m (800 ft)
120 m (400 ft)
1.50 mm
2
(AWG 15)
300 m (1
000 ft)
180 m (600 ft)
2.50 mm
2
(AWG 13)
300 m (1
000 ft)
300 m (1
000 ft)
5.1.4
Terminal assignment
Transmitter: supply voltage, input/outputs
The terminal assignment of the inputs and outputs depends on the individual order version of
the device. The device-specific terminal assignment is documented on an adhesive label in the
terminal cover.
Supply voltage
Input/output
1
Input/output
2
Input/output
3
Input/output
4
1 (+)
2 (–)
26 (A)
27 (B)
24 (+)
25 (–)
22 (+)
23 (–)
20 (+)
21 (–)
Device-specific terminal assignment: adhesive label in terminal cover.
Transmitter and sensor connection housing: connecting cable
The sensor and transmitter, which are mounted in separate locations, are interconnected by a
connecting cable. The cable is connected via the sensor connection housing and the
transmitter housing.
Terminal assignment and connection of the connecting cable → 19.