Gamma Modulator FHG65 Synchronizer FHG66
Hauser
5
18...3
VDC
5
1
1
4
2
3
A0018540
1
FHG65
2
Electrical connection between FHG66 and FHG65 (1)
3
Electrical connection between FHG66 and FHG65 (2)
4
FHG66
It is recommended to install the switch for the supply voltage near the device and to mark it as
a disconnector for the device.
Use of the Synchronizer FHG66, and particularly of its alarm output, is recommended for
minimum point level detection, as an undetected failure of the Modulator FHG65 may result in
faulty switching behavior
1
1
1
1
2
2
A
B
A0021133
A
Maximum point level detection
B
Minimum point level detection
1
FHG65
2
FHG66
Cascading multiple
Synchronizers FHG66
If more than three radiation sources are used, the synchronization chain must be extended by
cascading: here, an additional synchronizer (3) is connected to one of the outputs of the
synchronizer (2) instead of a modulator. All connected Gamma Modulators then operate in common