Operation
24
• Enter receiver's telephone number.
• OK
• Now enter the telephone number under which the unit connected by modem can be reached and
start the connection with "OK".
Fig. 14:
Modem connection at unit and PC
5.6.5
Communication via Ethernet (TCP/IP)
In principle, all units equipped with an internal Ethernet interface can be integrated into a PC
network (TCP/IP Ethernet).
The unit(s) can be accessed by any of the PCs in the network using the provided PC software. There
is no need to install driver software ("COM redirection") on the PC because the provided PC
software has direct access to the Ethernet.
The system parameters "IP address", "Subnetmask" and "Gateway" are input directly at the unit.
Changes to the system parameters are not activated until the SETUP menu is closed and the settings
accepted. Only then will the unit work with the new settings.
!
Note!
It is not possible for several clients (PC) to communicate with a server (unit) at the same time. If a
second client (PC) tries to establish a connection, he receives an error message.
Ethernet commissioning
Before a connection can be established via the PC network, the system parameters in the unit must
be set via "Setup - Communication - Ethernet".
!
Note!
You can obtain the system parameters from your relevant network administrator.
The following system parameters must be set:
1.
IP address
2.
Subnetmask
3.
Gateway
!
Note!
This menu does not appear unless the unit is equipped with an internal Ethernet interface.