![Endress+Hauser Deltabar S FMD 76 Скачать руководство пользователя страница 5](http://html1.mh-extra.com/html/endress-hauser/deltabar-s-fmd-76/deltabar-s-fmd-76_technical-information_2407113005.webp)
Deltabar S
Hauser
5
Measuring principle
Ceramic process isolating diaphragms used for PMD70 and FMD76
The ceramic measuring cell is based on the principle of a plate capacitor with an electrode on (1) and a movable
electrode on the interior of the diaphragm (3). Standard silicone oil or mineral oil filling oils for this measuring
cell.
A differential pressure (p
1
p
2
) causes a corresponding deflection of both diaphragms. Both capacitance values
are converted and are fed to the microprocessor of the transmitter as a digital signal.
Advantages:
• Self-monitoring for process isolating diaphragm break or oil loss (constant comparison of the measured
temperature with a temperature calculated from the capacitance values)
• Extremely high resistance to aggressive media
• Suitable for vacuums up to 1 mbar
abs
• Metal-free versions available
• Second process barrier (Secondary Containment) for enhanced integrity
Metallic process isolating diaphragms used for PMD75, FMD77 and FMD78
The process isolating diaphragms (3/9) are deflected on both sides by the acting pressures. A filling oil (4/8)
transfers the pressure to a resistance circuit bridge (semi-conductor technology). The differential-pressure-
dependent change of the bridge output voltage is measured and further processed.
Advantages:
• Standard operating pressures: 160 bar and 420 bar
• High long-term stability
• Very high single-sided overload resistance
• Second process barrier (Secondary Containment) for enhanced integrity
Ceramic process isolating diaphragms used for
PMD70 and FMD76
Metallic process isolating diaphragms used for PMD75, FMD77 and FMD78
P01-xMD7xxxx-03-xx-xx-xx-000
Ceramic measuring cell PMD70 and FMD76
1
Meter body
2
Process isolating diaphragm
3
Electrodes
4
Glass frit fixes the process isolating
diaphragm onto the meter body
5
Temperature sensor
P01-xMD7xxxx-03-xx-xx-xx-002
Metal measuring cell 10 mbar and 30 mbar
1
Sensing element
2
Silicon diaphragm
3
Process isolating diaphragm
4
Filling oil
5
Overload diaphragm
P01-xMD7xxxx-03-xx-xx-xx-003
Metal measuring cell as of 100 mbar
6
Sensing element
7
Overload diaphragm
8
Filling oil
9
Process isolating diaphragm
➀
➁
➃
➂
➄
1
p
2
p
➂
➃
➄
➀
➁
1
p
2
p
1
p
2
p
➆
➇
➈
➅