Maintenance
Cleanfit CPA875
30
Hauser
5.4.3
Molded seal - process connection, double chamber (C)
a0022299
Fig. 27:
Replacing seals, Part 1
1
Securing screws
1. Release the four securing screws (
pos. 1).
a0022300
Fig. 28:
Replacing seals, Part 2
2
Process connection
3
Gasket
4
Molded seal
8
Immersion tube
2. Remove the process connection (pos. 2).
3. Remove the molded seal (pos. 4) from the
process connection
4. Apply a thin layer of grease to the new
molded seal (e.g. Klüber Paraliq GTE 703).
5. Slide the molded seal over the immersion
tube (pos. 8) and into the guide groove of
the service chamber. Ensure that the
molded seal is seated correctly.
6. Position the gasket (pos. 3) on the rinse
chamber.
7. Place the process connection on the service
chamber.
8. Tighten the four securing screws with a
torque of 4 Nm.
Содержание Cleanfit CPA875
Страница 43: ...Cleanfit CPA875 Endress Hauser 43...
Страница 44: ...www addresses endress com 71237906...