CleanFit CPA 451
Installation
Hauser
11
3.3.4
Sensor installation
1.
Loosen the fastening screws (Pos. 1) and place them somewhere within reach (and
safely).
2.
Loosen the bayonet joint (Fig. 12) and pull the outer sleeve (pos. 7) including the
sensor holder (pos. 8) up as far as possible by the handles.
3.
Close the ball valve! To do this, press the hand lever (pos. 5) down as far as
possible (only possible in one direction!).
When the ball valve is closed the assembly is sealed off from the process.
4.
Pull the locking pin (pos. 4) out of the bayonet nut (pos. 6).
5.
Unscrew the bayonet nut (pos. 6) from the sensor holder (pos. 8) by holding the
outer sleeve (pos. 7) in place and turning the handles (pos. 2) anticlockwise.
6.
!
Hinweis!
Do not pull at the sensor cable
, when replacing the sensor! Use the bracket for
pulling out the sensor holder.
Pull the sensor holder completely out of the outer sleeve.
C07-CXA451xx-11-07-06-xx-001.eps
Fig. 11:
1
Fastening screws
C07-CXA451xx-11-07-06-xx-002.eps
Fig. 12:
2
Handles
3
Bayonet joint
4
Locking pin
5
Hand lever
C07-CXA451xx-11-07-06-xx-003.eps
Fig. 13:
6
Bayonet nut
7
Outer sleeve
C07-CXA451xx-11-07-06-xx-004.eps
Fig. 14:
8
Sensor holder
1
2
3
4
5
6
7
8