Vacuum principle
+
=
=
VP
D
-
3 The sample is now dosed to the
preset volume (VP). This is
dependent on the dosing tube
position (D). Excess liquid flows
back to the sampling point due
to a syphonic effect.
2 A fresh sample is then sucked
into the dosing chamber until
the conductivity level switch is
activated (sensors in the dosing
chamber flange)
4 The hose clamp is released and
the sample flows into the
composite container or bottles if
operating using sample
distribution.
1 The dosing system is
pneumatically sealed at the
beginning of each sample cycle.
The diaphragm pump then
blows the suction hose free of
obstructions via the dosing
chamber
asp-station d 2
Sampling principle
13