For sample lift less than 2 m we recommend using a 15 mm suction hose.
Both 13 mm and 15 mm connections are delivered.
The minimum conductivity of the sample liquid must not be less than
30 microsiemens !
•
The 13 mm internal
∅
hose must be of a spirally reinforced type.
•
The hose can be connected to the connector on the top left hand side of
the cabinet.
•
Maximum height difference: 6m from sampling point to sampler.
•
Maximum hose length: 30m
Rinse water pipe connection
Rinse water connection can be found inside the unit housing behind the left
hand container, for access both containers must be removed. External
pipework can be laid as required. Pipe size is 3/4", ID 19 mm. The minimum
water pressure must not be below 1 bar and the maximum water pressure
must not exceed 4 bar.
asp-port d 2 se
Mechanical installation