Outputs
Alarms and signals are retransmitted using the three outputs. These outputs
are constructed as open collectors and can be individually allocated in
setting up addresses 150 to 152.
Transistor Outputs 1,2 and 3 is dependent on the settings “standard” or
“inverse” in address 150 to 152.
Standard:
Open in "Alarm" and "Power off" and conduct in "Normal
operation".
Inverse:
Open in "Normal operation" and "Power off" and conducting
on "Alarm".
Power supply
Mains version (AC)
Plug in the mains plug (female) on the mains cable to the controller socket C
(male) and tighten the securing nut. Pull out rubber cable entry in housing
wall and lay cable inside
Plug connections:
Numbers are engraved on the plug
1= Black(L),
2 = free,
3 = Blue(N)
= Green/yellow (PE)
asp-port d 2 se
Mechanical installation
9